Luminex 200 System User Manual
18
7. Verify that the sample probe has been vertically aligned for the
plate used in the kit.
Adjust the sample
probe vertical
height
Adjust the sample probe vertical height each time you change the
type or style of microtiter plate.
1. Remove the clear plastic shield that covers the sample probe
area.
2. In a 96-well microtiter plate where overall height is no more than
0.75 inches (19 mm), place the appropriate alignment tool in the
plate:
For a standard plate with flat-bottom wells
—stack two of the
larger (5.08 mm diameter) alignment discs together and place
them into the selected well.
For a filter bottom plate
—stack three of the larger (5.08 mm
diameter) alignment discs together and place them into the
selected well.
For a half-volume plate with flat-bottom wells
—stack two of
the smaller (3.35 mm diameter) alignment discs together and
place them into the selected well.
For a round-bottom (U-bottom) plate
—stack two of the
smaller (3.35 mm diameter) alignment discs in the selected well.
For a plate with conical wells
—place one alignment sphere into
the selected well.
3. Eject the plate holder. Place the 96-well microtiter plate on the
Luminex XYP instrument plate holder with position A1 in the
top left corner.
4. Verify that the correct well location is selected in the Luminex
software, and that you are using the appropriate number of
alignment discs. Retract the plate.
5. Loosen the front thumbscrew on the probe holder one-third to
one-half turn. Pull it upward until it touches the top of the
adjustment slide. Tighten the thumbscrew.
6. Use the Luminex software to lower the sample probe.
7. Loosen the front thumbscrew. Gently push the probe down until
it just touches the top of the alignment discs or sphere.
8. Tighten the front thumbscrew.
Note:
Verify that the
microtiter plate is not
warped. Warped plates can
lead to incorrect probe
height adjustment.
Note:
Alignment discs can be
placed in any well as long as
the well is designated in the
software.
Содержание Luminex 200
Страница 1: ...Luminex 200 System User Manual Lum nex...
Страница 9: ...iii Luminex 200 System User Manual Hardware 47 Software 48 xMAP Reagents 49...
Страница 10: ......
Страница 42: ......
Страница 56: ...Luminex 200 System User Manual 46...
Страница 60: ...Luminex 200 System User Manual 50...
Страница 64: ......