Service manual
Service Manual – 7 / 63
XE_ABS_MT_AN_200224
5.
ULTRASOUND SYSTEM PROBES SPECIFICATIONS
5.1.
Standardized A probe
Probe Reference
STD-A
Frequency
8 MHz
Focal Length
Non focused
Emission running mode
Pulsed
Emission Repetition
Rate
67 Hz
Axial resolution
0.2 mm (at - 6dB)
5.2.
Biometry probe
Probe Reference
TP-01-b (Tono-Probe)
or TP-02-las (Probeam)
Frequency
11 MHz
Focal Length
20 to 25 mm
Emission running mode
Pulsed
Emission Repetition
Rate
67 Hz
Active diameter
5 mm
Active surface
20 mm2
Axial resolution
0.15 mm (at - 6dB)
5.3.
15 MHZ B-SCAN probe
Probe Reference
B1
Sector angle
50°
Frequency
15 MHz
Emission-running
mode
Pulsed
Emission Repetition
Rate
2720 Hz
Focal length
23mm to 25mm
Active diameter
7.5mm
Active surface
44mm²
Axial resolution
115µm (at -6dB)
Lateral resolution
400µm (at -6dB)