Страница 1: ...pe er ra at to or rs s M Ma an nu ua al l GF0500067REVA05 LowAirLossTherapy MattressReplacementSystem Model 763400A P Pl le ea as se e r re ea ad d t th he e m ma an nu ua al l b be ef fo or re e u us...

Страница 2: ...Contents Important Safeguards 2 Introduction 3 Product Description 4 Installation 5 Operation 7 Cleaning 10 Storage 11 Maintenance 11 Troubleshooting 12 Technical Description 13...

Страница 3: ...ice center for examination and repair 5 Keep the cord away from heated surfaces 6 Never block the air openings of this product or place it on a soft surface such as a bed or couch where their openings...

Страница 4: ...61000 4 2 IEC 61000 4 3 IEC 61000 4 4 IEC 61000 4 5 IEC 61000 4 6 IEC 61000 4 8 IEC 61000 4 11 EN55011 CLASS B EN 61000 3 2 EN 61000 3 3 FOR US AND CANADA ONLY 1 2 Intended Use The system is a true l...

Страница 5: ...It also provides excellent ventilation and keeps patients dry from humidity The 10 inch high mattress which consists of 8 inch high air bladders at the top and a layer of 2 foam at the bottom provide...

Страница 6: ...bed frame Please note for the foot end Mattress can be secured firmly by fixing the elastic straps at the bottom bed frame NOTE Please cover the mattress with a cotton sheet to avoid direct skin cont...

Страница 7: ...er Do not use it for any other purpose 5 Then turn the main power switch ON The power switch is located on the right side panel of the pump unit NOTE Ensure the CPR connectors are sealed completely In...

Страница 8: ...r outage during normal use ALARM MUTE Enable and disable the audible alarm feature Press this button to silence the audible alarm The alarm buzzer will sound again after 3 minutes if the pressure has...

Страница 9: ...tomatically STATIC For true low air loss therapy mode similar to a regular air mattress with very low air pressure support and true low air loss therapy PULSATION For pulsating and true low air loss t...

Страница 10: ...re Indicator is off it indicates the mattress has reached to its preset softness Patient can lie on the mattress NOTE Check to see if the suitable comfort range or pressure setting is selected by slid...

Страница 11: ...button on the control panel When power failure situation occurs alarm will last until backup battery is low or it can be turned off by manual operation Once power failure alarm situation is resolved i...

Страница 12: ...oblems or malfunctions to local distributors 7 2 Air Filter Replacement The air filter should be checked and dusted periodically once per month 1 Two foam type filters are used One is located on the b...

Страница 13: ...oo long Release some air by disconnecting the air tube connector and change to preferred setting Air Mattress is loose Check if all the snap buttons or straps of mattress are all securely fastened Che...

Страница 14: ...orage 10 to 90 non condensing Shipping 10 to 90 non densing Atmosphere 700 hPa to 1060 hPa Classification Class II Type BF IPX0 Applied part Air Mattress Not suitable for use in the presence of a flam...