manualshive.com logo in svg
background image

Lumencor Light Engine Manual 

SPECTRA FIA 
Light Engine

SPECTRA FIA Light Engine Manual 57-10008

1

lumencor.com

Содержание 90-10176

Страница 1: ...Lumencor Light Engine Manual SPECTRA FIA Light Engine SPECTRA FIA Light Engine Manual 57 10008 1 lumencor com ...

Страница 2: ...ow Voltage Directive 2014 35 EU EMC Directive 2014 30 EU RoHS Directive 2011 65 EU 2015 863 EU REACH Regulation EC No 1907 2006 EC EU Declarations of Conformity can be found at https lumencor com company compliance Revision History Lumencor Inc 14940 NW Greenbrier Parkway Beaverton OR 97006 USA 503 213 4269 www lumencor com Revision C 011822 SPECTRA FIA Light Engine Manual 57 10008 1 lumencor com ...

Страница 3: ... 3 Installation and Operating Instructions 4 Spectral Distribution and Power Output 5 Temporal Response 6 Operational Specifications 7 Routine Maintenance and Trouble Shooting 8 Customer Support 9 Warranty SPECTRA FIA Light Engine Manual 57 10008 2 lumencor com ...

Страница 4: ...cross the heat sink fins and exhaust the heated air out the rear panel The light sources within the Spectra FIA light engine are controlled by software via one of two serial interfaces RS 232 or TCP to the onboard computer operating Lumencor firmware The user can enable or disable each source independently by serial commands as well as change the intensity of each source independently The only man...

Страница 5: ...e matériel Safety Items Mesures de sécurité Warning DO NOT use an unapproved power supply An external power supply is required for use with the Spectra FIA light engine The power supply used must be capable of supplying 24VDC 14A The equipment is required to be supplied by an approved certified DC power source meeting the minimum electrical ratings of the Spectra FIA Light Engine Avertissement NE ...

Страница 6: ...pièce réparable et l ouverture de son boîtier a pour effet d annuler la garantie Caution DO NOT set liquids on the light engine Spilled liquids may damage your light engine Attention NE PAS placer de liquide sur la source lumineuse Les liquides renversés peuvent endommager la source lumineuse Caution DO NOT drop the light engine It contains glass optical components that could be damaged or misalign...

Страница 7: ...air intake and exhaust Also shown are the locations of the light output ports A and B and the power supply and communications port connectors 3 2 Installation NOTE Any end product system incorporating or coupled to a Lumencor Light Engine shall be fully evaluated to verify all applicable safety and regulatory compliance requirements prior to use Spectra FIA Model Light Engine Part Number DC Power ...

Страница 8: ...m3 hour with a temperature 6 C above ambient Connect the DC power supply to the power input port front panel lower left corner using the DC power cable Lumencor part number 29 10098 or 29 10103 See Figure 2 for pin connections Position the unit in such an orientation that allows unrestricted access to the DC power connector In an emergency you may ned to disconnect power to the unit quickly Connec...

Страница 9: ...be controlled by a Lumencor GUI or via the built in library of software commands A complete listing of the commands is provided in the Lumencor Light Engine Command Reference Document Number 57 10018 The software commands can be transmitted via the RS 232 port or the LAN input The default IP address for LAN communications is 192 168 2 1 The current IP address can be determined using the GET IP com...

Страница 10: ...onger peak normalized to a value of 1 0 SPECTRA FIA Light Engine Manual 57 10008 9 lumencor com 400 500 600 700 800 900 Wavelength nm 0 5 10 15 20 Intensity mW nm SN8908 Output A Spectra Cyan Green Red NIR I NIR II Figure 5 Representative spectral output of Spectra FIA light engine Color Output A Power mW Output B Power mW Cyan 370 37 Green 886 88 Red 210 21 NIR I 188 19 NIR II 208 21 TOTAL 1862 1...

Страница 11: ...ope traces in Figure 6 show the initial trigger of the TCP pulse which is followed by the light pulse turning on or turning off SPECTRA FIA Light Engine Manual 57 10008 10 lumencor com Figure 6 Temporal response to TCP commands The oscilloscope trace marked in green represents the light output detected by an analog photodiode ...

Страница 12: ... Lifetime Designed for 20 000 hours light output Contains no components with MTBF less than 7 years Light Output Control Output light intensity commands The light sources are turned OFF when the output light intensity value is set to 0 Command examples The light sources are turned off SET MULCHINT 0 0 0 0 0 Command for 50 output light intensity SET MULCHINT 2048 2048 2048 2048 2048 Command for 100...

Страница 13: ...ering performance that is substantially at variance with the levels documented in the certificate of conformance The light engine must also have been used and maintained under operating conditions consistent with the specifications given in Section 6 and observing all the Precautions and Warnings notified in Section 2 This warranty does not extend to light engines that have been subject to misuse ...