INFRATHERM 

pyrometer IS 50-LO/GL 

 

 

21

 

 

Note:

 

The software program serves only as a means of setting the parameters of the signal 
processor and was not designed as a controlling mechanism for measurement systems! 
After set-up is complete, disconnect the PC adapter cable and close the converter 
housing. 

 
 

8.5.1. Connection 

PC / signal processor 

The program should only be used to set the parameters. Connect the PC and the converter in the following manner: 
 

• 

Disconnect the signal processor from the power supply.  

• 

Open the cover of the signal processor 

• 

Plug the PC adapter cable into the 6 pole socket and connect the other end to an available COM port interface on 
your PC 

• 

Reconnect the signal processor to the power supply  

• 

Run the PC program I50_SERV.EXE from the supplied diskette or CD 

 
 

 

 

Attention:

 

Prior to connecting:

 When the PC adapter cable is connected, a direct connection 

between the PC and the pyrometer is established. However, there is still the possibility 
that a voltage difference between the PC and the signal processor exists in this connec-
tion, which can impact the 4 - 20 mA current output of the pyrometer. This is especially 
important if the temperature control of your system (i.e. oven or furnace temperature) is 
dependent on the output signal from the pyrometer. 

 
 

8.5.2. The 

program 

window 

 

 

 
 
 
 
 
 
 
 
 

• 

Click on a flag to select your language. 

• 

Enter a lower and upper value for the temperature sub range or shift the controller accordingly. 

• 

Typically, it is only necessary to enter the emissivity for older models that are not capable of being set on the device 
itself. 

• 

By clicking on the “°C” button, the display is switched over to “°F“.  Clicking the button again the display switches 
back to “°C”.  

• 

You may display a digit after the decimal point by clicking on the “0.” button 

• 

Ti specifies the internal temperature of the converter 

• 

To exit the set-up screen, click on ”Exit.” 

• 

After the settings have been configured, immediately unplug the PC adapter cable from the signal processor and the 
PC and securely close the signal processor! 

 
 
 
 

Language selection 

Controller for  
beginning of  
sub range 

Controller for  
end of  
sub range 

Entering end of  
sub range 

Entering beginning 
of sub range 

Pyrometer data 

Selection   
°C / °F 

Display of a decimal 
place 

Current  
output current 

Current measuring 
temperature  

Internal temperature 
of converter 

Emissivity (only active at 
older  models) 

Содержание Impac IS 50-LO/GL

Страница 1: ...IS 50 LO GL INFRATHERM Pyrometer Betriebsanleitung User Manual IMPAC Spezialist f r ber hrungslose Temperaturmessung IMPAC Specialist in non contact thermometry www imaxsun com...

Страница 2: ...of additional analyzing devices 17 6 Mechanical installation 17 6 1 Overview 17 6 2 Converter 18 6 3 Fibre optical head 18 6 3 1 Ambient temperature 18 6 3 2 Serial number 18 6 4 Minimum bending radiu...

Страница 3: ...C Infrared GmbH have a warranty of two years from the invoice date This warranty covers manufacturing defects and faults which arise during operation only if they are the result of defects caused by I...

Страница 4: ...e must be within the limits of the full temperature range and must have a minimum span of 51 C Emissivity 5 100 set via emissivity control inside the pyrometer Response time 1 s set ex works Accuracy...

Страница 5: ...in a permanent readable condition 4 1 General Each person working with the pyrometer must have read and understood the user manual before operation Also this has to be done if the person already used...

Страница 6: ...n side of the power supply switch board the shield must be open to avoid ground loops 5 1 Connector assignment for the converter Connector 3 pole flange plug series 723 Franz Binder GmbH Co Germany Pi...

Страница 7: ...ble The fibre optic connectors are equipped with a special sealing ring to protect the connection against penetra tion of liquids Attention The light guide end of the fibre optic cable as well as the...

Страница 8: ...on from the measured object which passes through the lens and is then converted to an electrical signal The measured object may be at any distance from the pyrometer However the farther the measured o...

Страница 9: ...erature The emissivity setting of the pyrometer has to be adjusted accordingly Typical emissivity values are listed below Measuring object Emissivity at 0 8 1 1 m molten glass 96 Set up examples at th...

Страница 10: ...e 4 20 mA current output of the pyrometer This is especially important if the temperature control of your system i e oven or furnace temperature is dependent on the output signal from the pyrometer 8...

Страница 11: ...al head 11 2 Accessories 3 820 650 Connection cable length 5 m 3 820 660 Connection cable length 10 m 3 820 670 Connection cable length 15 m 3 820 680 Connection cable length 30 m On request Other len...

Отзывы: