IMPAC pyrometers IS 210 · IGA 210

 

 

 

 

10 

 

 

7.4 

Temperature sub range 

You have the opportunity to choose a sub range (minimum span 51°C) 
within the basic measuring range of the pyrometer. This sub range corre-
sponds to the analog output (4 to 20 mA). 
 
 

7.5 

Clear time t

CL

 for maximum value storage 

The maximum value storage stores the highest measurement value. The 
clear time defines the time period until the stored value is deleted and 
replaced by a new one.  
The following settings are possible: 

  Clear time Off: The storage is switched off and only momentary values 

are measured. 

  If any clear time is set the maximum value storage will be cleared at this time. 

  The “auto” mode is used for discontinuous measuring tasks. For example objects are 

transported on a conveyer belt and pass the measuring beam of the pyrometer only for a 
few seconds. Here the maximum value for each object has to be indicated. In this mode 
the maximum value is stored until a new hot object appears in the measuring beam. The 
temperature which has to be recognized as “hot“ is defined by the low limit of the adjusted 
sub range. The stored maximum value will be deleted when the temperature of the new 
hot object exceeds the low limit of the sub range by 1% or at least 2°C. 

 
 

8

 

Settings via service interface 

 

8.1 

with HT 6000

 (accessory)

 

The HT 6000 is a portable battery driven instru-
ment for the pyrometer parameter settings and 
indication of temperature. 
The pyrometer has to be connected to the HT 
6000 via its service interface cable with the fe-
male SUB-D connector. When the HT 6000 is 
switched on it recognizes the pyrometer auto-
matically and the preset parameters can be 
displayed. With the “

PAR

” button the different 

parameters can be called. With the arrow buttons 

 and 

 

the values of the parameters can be 

selected and confirmed with the “

ENTER

 

button. 

If no button is pressed for 30 s the HT 6000 
changes to the temperature indication 

without 

accepting the changed value

 

 

Display 

On/Off 

Selection of 
parameters 

Up / Down 

Confirmation 

SUB-D connector 

Settings: 

Min 

. . . 

Max 

Settings: 

Off 

50 ms 

. . . 

25 s 
auto 

Содержание IGA 210

Страница 1: ...IS 210 IGA 210 IMPAC Pyrometer Proven Quality Operation Manual Betriebsanleitung...

Страница 2: ...a USB adapter onto a PC 6 3 4 Connection to the portable parametrizing device HT 6000 7 4 Mechanical Installation accessories 7 5 Optics 7 6 Alignment with the LED targeting light 8 7 Parameter Settin...

Страница 3: ...s that arise from the use of any exam ples or processes mentioned in this manual or in case the content of this document should be incomplete or incorrect LumaSense Technologies reserves the right to...

Страница 4: ...peratures 75 C the output signal is set to approx 3 9 mA Load max 700 at 24 V power supply 400 18 V 1000 30 V Service interface for parameter programming with portable setup device HT 6000 or via USB...

Страница 5: ...dditional devices operating with this mains voltage e g transformers Touching mains voltage can be fatal An incorrect connection and mounting can cause serious health or material damages Only qualifie...

Страница 6: ...corder Digital display 230 V 4 20 mA white brown Schirm schwarz 24 V DC Power supply green yellow Power supply for LED targeting light White 24 V Pin 1 Brown 0 V Pin 2 Black Shield Green LED Pin 5 Yel...

Страница 7: ...least as big as the spot size of the pyrometer in the meas uring distance The following drawings show the pyrometer s spot sizes M in different measuring distances a The spot size M in a distance of a...

Страница 8: ...ch are different from the nominal distance of the selected optics may be calculated using the following formula D D M a a M 2 2 D D M a a M 1 1 6 Alignment with the LED targeting light For exact measu...

Страница 9: ...r the two spectral ranges of the in struments are listed in the emissivity table below The tolerance of the emissivity values for each material is mainly dependent on the surface conditions Rough surf...

Страница 10: ...until a new hot object appears in the measuring beam The temperature which has to be recognized as hot is defined by the low limit of the adjusted sub range The stored maximum value will be deleted wh...

Страница 11: ...surement curves Subsequent analysis graphically or in table form for print out or export Spot size calculator to calculate spot sizes in any distances 9 Transport packaging storage With faulty shippin...

Страница 12: ...suring distance see 5 Emissivity set too high Set lower correct emissivity corresponding to the material see 7 2 Lens contaminated Clean lens carefully see 10 2 Temperature indication too high Emissiv...

Страница 13: ...60 Connection cable 5 m 3 821 770 Connection cable 10 m 3 821 780 Connection cable 15 m 3 821 790 Connection cable 20 m 3 821 800 Connection cable 25 m 3 821 810 Connection cable 30 m 3 826 500 Portab...