Alarm function

In IR DP mode the alarm function is activated automatically. Alarm
triggering and intensity are defined by the upper and lower alarm
threshold.
These two limit values can be calculated from the measured dew
point temperature (TdP) and the threshold values individually
determined in configuration mode: upper threshold (Hi) and lower
threshold (Lo).
The sum of dew point temperature and upper threshold value (Hi)
constitutes the upper alarm limit value (TdP + Hi). 
The difference of dew point temperature (TdP) and lower
threshold value (Lo) constitutes the lower alarm limit
value (TdP - Lo).
If the surface temperature falls below the upper alarm
limit (TdP + Hi) an acoustic and visual alarm (flashing laser and
flashing ALARM display) is triggered which intensifies in
proportion to the increased approximation to lower alarm limit.
The more the surface temperature drops, the faster increases the
repetition frequency of the acoustic and visual signals. Maximum
repetition frequency is triggered when reaching the lower alarm
limit (TdP - Lo).

Example:

 The current dew point temperature (TdP) is +10 °C.

You define the upper threshold (Hi) as 5 °C and lower
threshold (Lo) as 5 °C. Then the alarm starts at +15 °C (TdP + Hi)
and reaches maximum intensity at 5 °C (TdP - Lo).

Temperature display configuration

The temperature can be displayed in either Celsius (°C) or
Fahrenheit (°F).
The temperature can be displayed as air temperature or surface
temperature.

1. Press  the  Up  key (9)  to  switch  between  °C  and  °F  in  the

respective operating modes.

– The temperature will be displayed in the selected unit in the

upper measured value display (16).

– The unit (17) will be indicated on the display (1).

Operating 
mode

Temperature unit

Description

TH

°C

Air temperature in °C

°F

Air temperature in °F

IR

°C

Surface temperature in °C

°F

Surface temperature in °F

IR DP

°C

Surface temperature in °C

°F

Surface temperature in °F

%WT

CM

m/s

ppm °F °C

17

Содержание XC250

Страница 1: ...p p p p p pou ou ou ou ou ou ou u ou ou ou ur r r r r r r r r r la la la la la la la la la la p p p p p p p p p p pr r r r r r r r r r ci ci ci ci ci ci ci ci ci cisi si si si si si si si si ion on on...

Страница 2: ...results Lufft Warranty The warranty is for 12 months Damages caused by incorrect use by untrained people or startup by unauthorised people are excluded from the warranty The device complies with the...

Страница 3: ...urface temperature of the measuring object and the dew point temperature within the measurement environment A minimum maximum and average function is available for the direct analysis of the measured...

Страница 4: ...Technical data Parameter Value Air temperature NTC Measuring range 20 to 50 C Tolerance 0 2 C 0 to 40 C otherwise 0 4 C Resolution 0 1 C Object temperature Pyrosensor Measuring range 70 to 380 C Toler...

Страница 5: ...Personnel qualifications People who use this device must have read and understood the operating manual especially the Safety chapter For maintenance or repair work which requires the housing to be ope...

Страница 6: ...bag to transport the device safely Storage When the device is not being used observe the following storage conditions dry protected from dust and direct sunlight with a plastic cover to protect it fr...

Страница 7: ...device is ready for operation The displayed units are based on the settings of the last utilization 4 Select the desired measuring mode 5 Point the device at the area to be measured The measured value...

Страница 8: ...measured value display 18 indicates either the humidity or the dew point To set the unit see Humidity display configuration on page 10 IR DP mode Risk from laser radiation Class 2 laser radiation Las...

Страница 9: ...l 2 2 Confirm with the OK button 12 CAL1 IR mode Specifying the temperature offset value range 10 to 10 C or 18 to 18 F Set the degree of emission value range 0 1 to 1 0 1 Enter the temperature offset...

Страница 10: ...e more the surface temperature drops the faster increases the repetition frequency of the acoustic and visual signals Maximum repetition frequency is triggered when reaching the lower alarm limit TdP...

Страница 11: ...values cannot be saved on the device itself In order to save measured values the device has to be connected to a PC via a USB cable using the SmartGraph software 1 Briefly press the OK button 12 The...

Страница 12: ...adiate Many organic materials and surfaces have a degree of emission of approx 0 95 Metal surfaces or shiny materials have a lower degree of emission and therefore only yield inaccurate measured value...

Страница 13: ...isit the website www smartgraph3 de download 2 Double click the downloaded file 3 Follow the instructions of the installation wizard Starting the PC software 1 Start the SmartGraph3 software 2 Switch...

Страница 14: ...il of 27th January 2003 concerning old electrical and electronic equipment After the end of its use please dispose of this device in a manner appropriate to the relevant legal requirements In the Euro...

Страница 15: ...820 E Mason St A Santa Barbara CA 93103 Tel 01 919 556 0818 Fax 01 805 845 4275 E Mail sales lufftusainc com www lufft com Lufft China Shanghai Of ce Lufft Shanghai Measurement Control Technology Co...

Отзывы: