Model 334A Operations Manual
Ludlum Measurements, Inc
45
January 2021
Chronic & Acute Charts
The subsequent tabs are charts for the Acute and Chronic windows for both Alpha data. You
can follow the same instructions as above to refresh the template with new data from your unit.
For additional isotopes-of-interest, create copies of the tabs. These points are supplemental and
may help with instruction:
Because there are several data tabs, you can either select
Refresh All
from the first data
tab and choose the files one after the other
in order
, or select
Refresh
from each tab and
choose one file at a time.
Alpha Chronic Chart
Alpha Acute Chart
When formatting the
Y-axis
, the whole chart does not need to be in view, as the
minimum detectable line will likely be too high in the beginning to achieve good
readability of the rest. The chart is best read when the typical readings are within the
bottom fifth of the chart. See the example chart below for typical appearance. Once
formatted, leave the axis set for better comparisons between data files.
The
X-axis
may also need to be formatted for different dose measurement lengths. The
length in time is automatic but the
major unit
needs to be fixed. For 1 hour increments
use 0.04166666, for 2 hours use 0.08333333, for 4 hours use 0.16666666, and for 8 hours
use 0.33333333.
Содержание 334A
Страница 1: ...LUDLUM MODEL 334A ALPHA AIR MONITOR January 2021 ...
Страница 2: ......
Страница 3: ...Model 334A Operations Manual ...
Страница 5: ......
Страница 7: ......