5.3. DATA LOG OPERATION
45
• The number will be blinking, indicating a changeable value. The user may then enter
the preferred Location ID Table index by a short press of the MODE (to increment the
value) as in Setup mode.
• Once the user has the preferred Location ID Table index entered, a short press of the
LOG button will save the log data.
• After displaying the Datalog Table index, the instrument will return to the previous
mode of operation.
• The Datalog Table is linear; once the table is full, several 1/4 second audio alerts will
sound and the maximum Datalog Table index (1.00 k) will be displayed for about 2
seconds, and no further writes will be allowed until the Datalog Table is erased.
5.3 Data Log Operation
•
LOG Button:
handle-mounted
•
Clock:
Real Time Clock
•
Location ID Storage:
internal storage of up to 250 32-byte location IDs
•
Datalog Record Storage:
internal storage of up to 1000 64-byte datalog records
5.4 Format
The datalogging format is as follows:
• Format Version (1 Byte)
• Month (1 Byte)
• Year (2 Bytes)
• Day (1 Byte)
• Hour (1 Byte)
• Minutes (1 Byte)
• Seconds (1 Byte)
• Alpha Logged Value (4 Bytes)
• Beta Logged Value (4 Bytes)
• Alpha Range (1 Byte)
• Beta Range (1 Byte)
December 2021