44
CHAPTER 5. DATALOGGING
• For approximately 2 seconds, the LCD will display the Datalog Table index for the
newly saved log data.
Figure 5.1: Displaying Datalog Table Index of 12
• After displaying the Datalog Table index, the instrument will return to the previous
mode of operation.
• The Datalog Table is linear; once the table is full, no further writes will be allowed until
the Datalog Table is erased.
• If an attempt is made to write to a full Datalog Table, several 1/4 second audio alerts
will sound and the maximum Datalog Table index (1.00 k) will be displayed for about
2 seconds. The instrument will then return to the previous mode of operation.
5.2 Datalogging Operation – Mode 2
Datalogging Mode 2 will allow the user to choose the Location ID (by Location ID Table in-
dex) to store with the logged data.
• When the LOG button is pressed, the current radiation readings and other log data are
temporarily stored.
• In COUNT Mode:
–
By default, Mode 2 is disabled during a count; instead, the datalogging feature
functions as it would in Mode 1.
• The LCD display will show a possible Location ID Table index for the user. The index
will be auto-incremented from the previously used index.
Figure 5.2: Displaying Datalog Location ID Table Index of 36
Ludlum Measurements, Inc.