4.4. SETUP MODE OPERATION
35
sired option.
- Normal:
All parameters are protected from change through the device menu. This is the
default mode when the Setup Protect dipswitch is configured to the ON position.
- Bypass Alarms:
All parameters except for Rate, Count, and Dose Alarms are protected from
change through the device menu.
- Bypass Alarms Count:
All parameters except for Rate Alarms, Count Alarms, Dose Alarms,
and the User Defined Scaler Count Time are protected from change through the device
menu.
- Bypass All:
All parameters may be changed from the Setup Menu, similar to leaving the
Setup Protect DIP Switch set to the OFF position.
ð
With the DIP Switch 2 in the ON position, setup mode may be entered through
the front panel, and any parameter which has not been bypassed may be
viewed, but changes cannot be made.
Only bypassed parameters may be
changed.
DISPLAY BACKLIGHT ‘Continuous On’:
The Model 3001 display backlight can be set to re-
main on continuously during operation. Follow the steps from
SETUP PROTECT
to access
the DIP switch but change DIP Switch 1 for display backlight selection. Setting DIP Switch 1
to the ON (forward) position will configure the display backlight to remain on during opera-
tion. Set DIP Switch 1 to the OFF (back) position, and the display will be backlit only when
light levels are low.
ð
Setting the display backlight for continuous-on operation can result in reduced
battery life. The backlight can be configured to always be off as well through the
device menu or Lumic 2.0 Calibration Software.
4.4 Setup Mode Operation
Once the Model 3001 is in Setup mode, the detector selection will be displayed on the LCD
and the detector number will be blinking indicating it as the selected item. Select the appro-
priate detector using the DETECTOR button. Once the detector is chosen, the setup page
selection will be displayed on the LCD and the Page number will be blinking, indicating it
as the selected item. The number of available parameters per Setup Page will be displayed
using the graph where the number of segments indicates the number of parameters. Short
press the MODE button to choose the Setup Page you are interested in. Once the desired
Setup Page is shown, short press the UNITS button to cycle down to the first parameter of
that Setup Page. Further short presses on the UNITS button will cycle down through the pa-
rameters for that page. Extra-long-pressing the UNITS button will go back up to the Setup
Page. Extra-long pressing again will return to the detector selection. Finally, another extra-
long press will exit the Setup Mode. To avoid all of the individual presses, press and hold
December 2021
Содержание 25009185
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 12: ......
Страница 14: ......
Страница 32: ......
Страница 35: ...29 Weight 0 7 kg 1 6 lb excluding detectors and clips December 2021 ...
Страница 36: ......
Страница 56: ......
Страница 62: ......
Страница 72: ......
Страница 74: ......
Страница 78: ......