Troubleshooting
5
Large residual amplitude
Check parameter file, do not fit
past point 170
Bad cables, especially PIO ribbon
cable
Bad grounding of fringe signal
10nm residual frequency swept
Wrong laser type selected
Incorrect laser modulation
frequency entered in the
fg5param.dat file
No fringes at all
No power to APD.
One/both beam blockers are
pushed in
Very bad alignment of the
test/reference beams. Look into
the telescope and make sure that
the test and reference beams are
overlapped. Also make sure that
the beams are overlapped in the
fringe viewer with the twiddler
Fringe amplitude varies as the cart is
moved up or down.
The iodine-stabilized laser is
running multi-mode. This can
indicate the laser servo gain being
set too high.
Table 5-2
Interferometer Problems and Solutions
5-7
Содержание Micro-g LaCoste FG5
Страница 7: ...Table of Figures This Page is Intentionally Blank vii ...
Страница 8: ......
Страница 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Страница 21: ...Design Components and Function 2 Figure 2 8 Side view of interferometer optics and beam path 2 11 ...
Страница 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Страница 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Страница 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Страница 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Страница 56: ......
Страница 59: ...Adjustment and Maintenance 4 4 3 ...
Страница 78: ...Adjustment and Maintenance 4 4 22 ...
Страница 106: ......
Страница 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Страница 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Страница 144: ......