How to Set Up and Run the FG5
3
3.1.1.2.WEO Model 100 Laser
•
Turn on the power (main power and HV key switches).
•
Select the proper iodine peak (usually “E”).
•
Set the servo control to AUTO.
•
Do not adjust any other controls.
Nominal control settings are:
Signal Monitor Section
:
Meter Select:
1F
Gain
1
Time Constant
1
Temperature Control Section:
Body Temp Mode
TEMP
Temperature Control Section
Meter
BIAS
Bias Voltage
Set to 0V (meter)
•
Allow the laser to warm up for at least two hours (or until the
temperature of the laser has stabilized)
before beginning observations.
3-3
Содержание Micro-g LaCoste FG5
Страница 7: ...Table of Figures This Page is Intentionally Blank vii ...
Страница 8: ......
Страница 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Страница 21: ...Design Components and Function 2 Figure 2 8 Side view of interferometer optics and beam path 2 11 ...
Страница 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Страница 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Страница 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Страница 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Страница 56: ......
Страница 59: ...Adjustment and Maintenance 4 4 3 ...
Страница 78: ...Adjustment and Maintenance 4 4 22 ...
Страница 106: ......
Страница 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Страница 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Страница 144: ......