Design: Components and Function
2
Figure 2-7
Top view of the service ring.
2.1.5.VIEWING PORT
The
viewing port
(Figure 2-2 and Figure 2-3) is located in the top flange of the
dropping chamber. It allows visual observation of the dropping chamber
interior when the rotation monitor is not fitted to the system. The rotation
monitor (when fitted to the system) is mounted to the top flange of the dropping
chamber, directly above the viewing port. When the rotation monitor is not
mounted, a cover for the port is used to exclude ambient light from the interior
of the dropping chamber during measurements.
2.1.6.THE DROP
In drop mode, a signal from the computer to the dropper controller initiates the
drop sequence. The cart drag-free chamber is driven slowly from its bottom
position to the “hold” position at the top of the drop. A second pulse initiates
the drop, and the cart accelerates downward at more than 1 g, leaving the test
mass in free-fall.
2-7
Содержание Micro-g LaCoste FG5
Страница 7: ...Table of Figures This Page is Intentionally Blank vii ...
Страница 8: ......
Страница 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Страница 21: ...Design Components and Function 2 Figure 2 8 Side view of interferometer optics and beam path 2 11 ...
Страница 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Страница 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Страница 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Страница 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Страница 56: ......
Страница 59: ...Adjustment and Maintenance 4 4 3 ...
Страница 78: ...Adjustment and Maintenance 4 4 22 ...
Страница 106: ......
Страница 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Страница 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Страница 144: ......