 alarm

Remote alarm case

EN

 Instructions for Use 

Содержание VENTIremote alarm

Страница 1: ...VENTIremote alarm Remote alarm case EN Instructions for Use...

Страница 2: ...arm case 7 3 2 Display film 9 4 Preparation and operation 10 4 1 Before operating for the first time 10 4 2 Setting up and connecting the device 12 4 3 Operation with the therapy device 14 4 4 Changin...

Страница 3: ...tents 10 Annex 21 10 1 Technical specifications 21 10 2 Safety distances 22 10 3 Markings and symbols 22 10 4 Standard scope of delivery 23 10 5 Accessories and spare parts 24 10 6 Warranty 24 10 7 De...

Страница 4: ...ing staff and relatives when looking after a ventilated patient in a domestic environment 1 2 Description of function VENTIremote alarm is connected to the device via a cable 10 m and 30 m long If the...

Страница 5: ...or use the VENTIremote alarm should be started up only once the temperature of the device is within the permitted range for operation Electromagnetic compatibility The product is subject to special pr...

Страница 6: ...ty information in these instructions Indicates an unusually significant hazardous situation If you ignore this instruction severe irreversible or fatal injuries may result Indicates a hazardous situat...

Страница 7: ...indicator This green LED indicates that the VENTIremote alarm is working correctly 3 Battery indicator If this yellow LED flashes the battery needs changing 4 Alarm indicator In combination with the b...

Страница 8: ...larms on to the VENTIremote alarm 3 HRS connector This connects the cable to the device 3 1 2 Cable for VENTIlogic LS VENTIlogic plus series prisma CHECK 1 ODU connector This connects the cable to the...

Страница 9: ...EN 9 Overview 3 2 Display film SYMBOL MEANING 1 Red LED on Alarm 2 Yellow LED flashing every six seconds Battery almost dead 3 Stand by indicator The green LED must light briefly every six seconds dur...

Страница 10: ...ration and operation 4 1 Before operating for the first time Before you can operate the VENTIremote alarm you must insert the U9VL BP type battery included in the scope of delivery 1 Open the battery...

Страница 11: ...ment 4 Push the battery compartment cover closed until it engages with an audible click The VENTIremote alarm is now ready for operation Notice Ensure that the connecting cables are underneath the bat...

Страница 12: ...remote alarm on a level surface 2 Connecting to LMT150TD Connect the ODU connector to the remote alarm connection of the device Warning Ensure that the VENTIremote alarm is free standing and not cover...

Страница 13: ...operation 3 Connecting to prisma CHECK Connect the bayonet connector to the remote alarm connection of the device 4 Connecting to VENTIlogic LS VENTIlogic plus series Connect the bayonet connector to...

Страница 14: ...every six seconds during operation to indicate operational readiness 4 3 2 Output of alarms If an alarm is issued on the therapy device the alarm indicator red LED of the VENTIremote alarm comes on pe...

Страница 15: ...anging the battery The VENTIremote alarm is operated with a battery of the U9VL BP type As soon as the battery indicator yellow LED comes on every six seconds and an acoustic alarm is issued every six...

Страница 16: ...es in the battery compartment 5 Push the battery compartment cover closed until it engages with an audible click Notice Ensure that the connecting cables are underneath the battery in the battery comp...

Страница 17: ...g of the VENTIremote alarm and the cable We recommend TERRALIN for this purpose See also the instructions for use for the disinfectant used We recommend using suitable gloves e g household or disposab...

Страница 18: ...see 6 2 2 Function check of the battery page 19 Function check of the entire system see 6 2 3 Function check of the entire system page 19 6 1 Intervals Perform a function check daily and before the d...

Страница 19: ...conds the battery needs replacing 6 2 3 Function check of the entire system To check the connection between the VENTIremote alarm and the device proceed as follows 1 Connect the VENTIremote alarm to t...

Страница 20: ...your local authority The device packaging cardboard and inserts can be disposed of in paper recycling facilities FAULT CAUSE OF FAULT REMEDY VENTIremote alarm does not switch on Battery discharged Ins...

Страница 21: ...rt and storage 5 C to 40 C 25 C to 70 C Permitted humidity for operation transport and storage Relative humidity 10 to 95 no condensation Signal volume 80 dB 4 dB A with a full battery at a distance o...

Страница 22: ...wer of the HF device in W Safety distance depending on transmission frequency in m 150 KHz 80 MHz 80 MHz 800 MHz 800 MHz 2 5 GHz 0 01 0 04 0 04 0 08 0 1 0 11 0 11 0 22 1 0 35 0 35 0 70 10 1 10 1 10 2...

Страница 23: ...e alarm 10 m packed for LM150TD LMT 31560 DESCRIPTION ORDER NUMBER VENTIremote alarm WM 27764 Cable 10 m VENTIremote alarm for LM150TD LMT 31595 Battery 9 V lithium WM 12166 Instructions for use for V...

Страница 24: ...ealer VENTIremote alarm 10 m packed for WM110TD WM120TD VENTIlogic LS VENTIlogic plus series WM 22745 DESCRIPTION ORDER NUMBER VENTIremote alarm WM 27764 Cable 10 m VENTIremote alarm for WM110TD WM120...

Страница 25: ...any the manufacturer of the devices described in these instructions for use hereby declares that the product complies with the relevant regulations of the Medical Device Directive EU 2017 745 The unab...

Страница 26: ......

Страница 27: ......

Страница 28: ...LMT 66735a 04 2021 EN L wenstein Medical Technology GmbH Co KG Kronsaalsweg 40 22525 Hamburg Germany T 49 40 54702 0 F 49 40 54702 461 www loewensteinmedical de...

Отзывы: