6 Hygiene treatment
6.3.2 Extended hygiene treatment on change of patient
1. Replace air filter, pollen filter, and bacteria filter.
2. Reprocess the device and components in accordance with the table below:
(see "6.6 Function check", page 35)
6.4 Clean air filter (gray filter)
1. Clean air filter under running water.
2. Allow air filter to dry.
PART
DISINFECTING
STERILIZATION
Housing including device outlet
port/inlet, power cord
Disinfect by wiping
(recommended products:
terralin
®
protect or perform
advanced Alcohol EP)
Not permitted
High-gloss surfaces on the
housing
Leakage circuit
Disinfect by immersion
(Recommended product:
gigasept FF
®
).
Rinse out circuit with clean water
and shake out thoroughly.
Dry circuit.
Not permitted
Single circuit with valve
Not suitable for reuse. Follow the
associated instructions for use.
Not permitted
Circuits for mouthpiece
ventilation
Heatable circuits
Follow the manufacturer’s instructions for use. Avoid
damage when using or reprocessing, especially to the
connecting cable and to the inner protective film over
the heated wire.
Mask
Follow the manufacturer’s instructions for use.