6 Hygiene treatment
6.3 Hygiene treatment for device
6.3.1 Cleaning device and components
1. Clean the device and components in accordance with the table below:
(see "6.6 Function check", page 35)
.
Risk of injury from electric shock!
Ingress of liquids may lead to a short-circuit, injure the user and damage the device.
Disconnect the device from the power supply before the hygiene treatment.
Do not immerse the device and components in liquids.
Do not pour liquids over the device and components.
Material damage as a result of ingress of liquids!
The device may be damaged by the ingress of liquids.
Use the circuit only when completely dry.
If you are using a heatable circuit or a circuit with an active exhalation valve, follow
the associated instructions for use.
PART
CLEANING
Housing including device
outlet port/inlet, power cord
Wipe down: Use water or mild detergent.
High-gloss surfaces on the
housing
Wipe down: Use water or mild detergent; do not use
microfiber cloth.
Leakage circuit
Rinse: Use hot water and mild detergent. Allow to dry
completely.
Single circuit with valve
Circuits for mouthpiece
ventilation
Heatable circuits
Follow the manufacturer’s instructions for use. Avoid
damage when using or reprocessing, especially to the
connecting cable and to the inner protective film over the
heated wire.
Mask
Follow the manufacturer’s instructions for use.