2 Safety
LM
T 6
86
51
c 0
2/2
02
2
EN
|
5
2.1.2 Electromagnetic compatibility
The device is subject to special precautions with regard to
EMC (electromagnetic compatibility). If these precautions
are not followed, the device may malfunction and individuals
may be injured.
Portable high-frequency communication equipment
(e.g. radios and cell phones), including their accessories
such as antenna cables and external antennas, for
example, must be used at a distance of at least 30 cm
from the device and its cables.
Do not use the device in the vicinity of active high-
frequency surgical equipment.
Operate the device within the EMC environment
specified for this device
interference immunity”, page 35)
in order to prevent
key performance characteristics being affected - for
example, ventilation parameters being affected by
electromagnetic interference.
Do not operate the device if the housing, cables or other
equipment for electromagnetic shielding are damaged.
The use of third-party accessories, third-party inverters,
and third-party cables may lead to increased
electromagnetic interference or reduced
electromagnetic interference immunity of the device
and to faulty operation. Only use the manufacturer’s
cables.
Do not operate the device in the immediate vicinity of
other devices or in a stacked arrangement, otherwise
there may be malfunctions. If it is necessary to operate
the device in the immediate vicinity of other devices or
in a stacked arrangement, keep all the devices under
observation to ensure that they are all operating
properly.
2.1.3 Energy supply
Operating the device outside the specified energy supply
may cause personal injury, damage the device or impair the
performance of the device.
Operate the power supply unit only at voltages from
100 V to 240 V.
Use DC cable LMT 31597 for operation on voltages of
12 V or 24 V.
Keep access to the power supply connector and the
power supply free at all times.
When using a battery-operated wheelchair: Connect
the device to the wheelchair battery only if a connection
of that kind is expressly provided in the instructions for
use for the wheelchair.
When operating using the cigarette lighter socket in a
car: Disable the car’s auto start/stop feature. Start the
car first, then connect the device.
2.1.4 Handling oxygen
Supplying oxygen without a special safety device can lead to
fire and injure people.
Follow the instructions for use for the oxygen supply
system.
Set up oxygen sources at a distance of over 1 m from
the device.
The oxygen rate supplied in l/min must not exceed the
oxygen flow prescribed by the medical professional.
The oxygen rate supplied in l/min may not exceed the
set HFT flow rate.
At the end of therapy, shut off the oxygen supply and
allow the device to run on briefly to flush residual
oxygen out of the device.
2.1.5 Transport
Operating the device in any kind of carrying bag may impair
device performance and injure the patient. Water and dirt in
the device may damage the device.
Only operate the device in the associated LUISA mobility
bag.
Transport or store the device in the associated LUISA
protective bag.
2.1.6 Wireless module
The device contains a wireless module. Operating the device
in the immediate vicinity of people and/or other antennas
may injure people, damage the device or impair device
performance.
Set up the device at least 20 cm away from any people.
Do not set up or operate the device with other
antennas.
2.2 General information
•
In order to react to an alarm and, if necessary, to use
emergency ventilation, you must subject both patient
and device to regular monitoring.
•
The use of third-party articles may lead to incompatibility
with the product. In such cases, please be aware that any
claim under warranty and liability will be void if original
spare parts are not used.
•
Connection by cable to a patient monitor is not a
substitute for a remote alarm system. Alarm data are
transmitted only for documentation purposes.
•
Have measures such as repairs, servicing, and
maintenance work, as well as modifications to the
product, carried out exclusively by the manufacturer or
by specialists expressly so authorized by the
manufacturer.
•
Connect only the licensed products and modules in
accordance with these instructions for use. The products
must meet the product standard applicable to them.
Non-medical equipment should be positioned out of the
patient's vicinity.