7 Alarms
24
|
EN
LM
T
68
65
1c
0
2/
20
22
7.4 Technical alarms
DISPLAY
CODE
CAUSE
ACTION
Service necessary. Please
get in touch with your
specialist dealer.
Various
Technical fault which can only be
eliminated by a specialist dealer.
Contact your specialist dealer. Have device
repaired.
Fault on touch display
173
Touch controller has failed.
Use on/off key to restart the device.
Intake air temperature
high
262
Ambient temperature too high.
Operate device at an ambient temperature of
5 °C to 40 °C.
Main board temperature
high
263
Ambient temperature too high.
Operate device at an ambient temperature of
5 °C to 40 °C.
Computer module
temperature high
264
Ambient temperature too high.
Operate device at an ambient temperature of
5 °C to 40 °C.
Unable to reach flow
364
Set flow not reached.
Check flow setting and accessories.
Disconnection device
outlet port
460
Circuit is not connected to the device
correctly or is not connected at all.
Check circuit and the correct fit of circuit.
Disconnection airway
pressure
461
Pressure measuring tube is not
connected to the device correctly or is
not connected at all.
Check pressure measuring tube.
Disconnection
exhalation module
463
Exhalation module is not connected to
the device correctly or is not connected
at all.
Check exhalation module.
Disconnection patient
464
Device operated with open patient/
ventilator interface (mask not applied).
Check circuit and patient/ventilator interface.
Double circuit selected in menu but
expiration tube not connected.
Double circuit selected in menu but
single circuit with valve or leakage
circuit connected.
Have the medical professional or the specialist
dealer set the connected circuit in the device.
Temperature of battery
E1 critically high
547
External battery 1 too warm.
Battery will switch off due to temperature.
Operate device at an ambient temperature of
5 °C to 40 °C.
Temperature of battery
E2 critically high
548
External battery 2 too warm.
Battery will switch off due to temperature.
Operate device at an ambient temperature of
5 °C to 40 °C.
Error internal battery
549
Internal battery defective.
Contact your specialist dealer. Have internal
battery replaced.
Battery capacity critical
550
Battery discharged (remaining battery
life: 5 minutes)
Connect the device to the power supply.
Battery capacity low
551
Battery discharged (remaining battery
life: 15 minutes)
Connect the device to the power supply.
No internal battery
553
No internal battery.
Contact your specialist dealer. Have internal
battery inserted.