4 Preparation and operation
4.1 Setting up and connecting
device
1. If required: Tilt the device to a horizontal or vertical
position.
The display adapts to the orientation automatically.
2. Connect the power cord to the power supply unit and
the socket.
3. Connect the power supply unit cable to the device.
4.2 Connecting circuit
Risk of injury from inadequate therapy if air inlet and
air outlet are blocked!
A blocked air inlet and/or air outlet can cause the device to
overheat, impair therapy, and damage the device.
Keep the filter compartment clear (
symbol).
Keep the device inlet port free (
symbol).
Keep the intake area of the cooling fan free
(
symbol).
Material damage from overheating!
Excessive temperatures may lead to the device overheating
and damage the device.
Do not cover device and power supply unit with textiles
(e.g. bedclothes).
Do not operate device in the vicinity of a radiator.
Do not expose device to direct sunlight.
Only operate the device in the associated mobility bag
for mobile use.
Alternatively, you can connect a direct voltage
electricity supply (12 VDC or 24 VDC) as per
ISO 80601-2-72.
Risk of asphyxia if invasive or non-invasive patient/
ventilator interfaces without an exhalation system are
used!
If invasive or non-invasive patient/ventilator interfaces
without an integrated exhalation system are used, CO
2
concentration may rise to critical values and put the patient
at risk.
Use invasive or non-invasive patient/ventilator interfaces
with an external exhalation system if there is no
integrated exhalation system.
Follow the instructions for use for the exhalation
system.
Risk of injury from potential patient disconnection!
On circuits without proximal pressure measurement and
with additional accessories such as HME or tube extension,
for example, it is not possible to detect patient disconnection
reliably.
Use the VTe low alarm for the double circuit and the
VTi high alarm for the single circuit with valve.