6 | Preparing the system for use
64
Basic Instructions for Use elisa 300-500
|
SW 2.09.x
|
Rev. 11
Remark
The batteries used have a service life of 500 charge cycles.
6.2
Gas supply
Connect the elisa 300 / elisa 500 O
2
gas inlet to an O
2
gas supply that meets the
requirements indicated in the technical specifications.
Remark
Technical details on gas supply can be found in chapter 12.3
“Supply”.
The current data of the connected gas supply can be reviewed any time during
ventilator operation.
Warning
Only use gas supply tubes that meet the local standards. A faulty
O
2
supply can pose a risk to the patient.
Warning
Ensure that only dry and clean gas is used. Water, oil or particles
entering the device can damage the device and endanger the pa-
tient.
If in doubt, provide suitable water traps and/or oil separators.
elisa 300 / elisa 500 has been designed to allow operation without O
2
supply at
any time. In this case, AIR will be supplied by the turbine integrated in the device.
If O
2
is not available for the system test, the absence of O
2
can be confirmed and
the system test can be carried out all the same. The test steps for the missing O
2
are skipped. After a system test without oxygen, the value for FiO
2
is automatically
set at 21%.
If FiO
2
is set to a value greater than 21% during the ventilation procedure, an
alarm will be displayed to indicate that O
2
is missing.
To view information about the available gases, press the “System” button and
select the “Gases” tab.
Remark
If the missing O
2
is connected during operation, the device auto-
matically switches to operation with oxygen.
The information that the system test was carried out without O
2
remains displayed until the system test is repeated.
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...