Functions | A 3
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
45
Warning
Observe the instructions for use supplied with the measurement
kit and the accessories used.
A 3.6.1
Connecting the intravesical pressure measuring kit
Connect the intravesical pressure measuring kit to the Paux II port of elisa 500.
Urine collection bag
Multi-gang manifold
Infusion
syringe
Saline infusion
Urinary
catheter
Warning
Check that the intravesical pressure measurement kit is connect-
ed to the correct Paux port. The intra-abdominal pressure cannot
be measured when the kit is connected to any other port; further-
more, this may jeopardise the safety of the patient.
A 3.6.2
IAP function
Remark
IAP is available to the end user only if it was enabled on the con
-
figuration level.
Open the IAP workspace by pressing the “Functions” button and then by select
-
ing the “IAP” tab. If configured accordingly, the intra-abdominal pressure measure
-
ment can also be selected via the shortcut “IAP”.
Activate and deactivate the function with the “On” and “Off” buttons.
IAP is displayed as a numeric value and as a graphic trend.
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...