Functions | A 3
Instructions for Use elisa 300-500 | Supplement A |
SW 2.09.x
| Rev. 11
37
Slip one of the two ends of the Y-piece onto the Paux I port of elisa 500.
The two ends of the Y-piece are identical, they are both closed with a non-return
valve. The valve allows gas to flow through in only one direction. Therefore, the
cuff pressure is maintained even if the Cuffscout line becomes disconnected.
Warning
Be sure to connect the Cuffscout line to
P
aux
I
. Connecting the
line to any other port may jeopardise the safety of the patient.
Warning
Cuffscout lines are single-use items which must be disposed of
after use. Use a new Cuffscout line for each new patient.
Remark
The maximum service life of the Cuffscout connecting line is 29
days.
A 3.4.2
Cuffscout function
You can choose between two modes: “
static
” and “
automatic
”.
At the configuration level, the cuff modes can be enabled and disabled and the pa
-
rameters for the automatic cuff mode can be defined according to the ward-spe
-
cific protocols.
You reach the Cuffscout menu by pressing the “Functions” button, followed by
selection of the “Cuffscout” tab.
The following parameters can be adjusted:
Cuff const.
Specified value for the static cuff pressure.
Cuff max
Safety window for the maximum cuff pressure or setting for
the maximum block manoeuvre.
Cuff min
Minimum cuff pressure
Remark
Initially, the cuff needs to be blocked manually.
Remark
You cannot switch off elisa 500 while the Cuffscout function is
active.
A 3.4.2.1
Cuff pressure in the standby mode
On completion of the ventilation procedure, the monitoring and control of the
cuff pressure can be continued, e.g., for tracheotomised patients. For this function,
Содержание elisa 300
Страница 3: ...elisa 300 elisa 500 Instructions for Use Basic Manual ...
Страница 10: ...8 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 0 Contents This page is intentionally left blank ...
Страница 18: ...2 Use 16 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 62: ...5 General information 60 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 ...
Страница 102: ...7 Operation 100 Basic Instructions for Use elisa 300 500 SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 217: ...elisa 300 elisa 500 Instructions for Use Supplement A Manoeuvres Functions Ventilation Modes ...
Страница 246: ...A 3 Functions 30 Instructions for Use elisa 300 500 Supplement A SW 2 09 x Rev 11 This page is intentionally left blank ...
Страница 351: ...CO2 Mainstream Sensor LeoCap Instructions for Use Supplement B ...
Страница 362: ...12 Instructions for Use Supplement B CO2 Sensor LeoCap Rev 04 elisa 300 elisa 500 elisa 600 elisa 800 ...
Страница 379: ...Multi Gas Sensor LeoLyzer Instructions for Use Supplement C ...
Страница 407: ...Pulse Oximetry and LeoClac Instructions for Use Supplement D ...
Страница 423: ...Instructions for Use Supplement D Pulse Oximetry and LeoClac Rev 01 17 This page is intentionally left blank ...
Страница 425: ...elisa 300 elisa 500 elisa 600 elisa 800 elisa 800 VIT External Communication Instructions for Use Supplement E ...
Страница 432: ...8 Instructions for Use Supplement E External Communication Rev 02 Rear view of elisa 600 Rear view of elisa 800 ...
Страница 437: ...Medical Equipment Interface elisa megs Instructions for Use Supplement F ...
Страница 440: ...4 Instructions for Use Supplement F elisa megs Rev 03 This page is intentionally left blank ...