8. Maintenance
Technical safety controls
Daily testing
The operator must visually inspect the humidifier to
check for the following:
▪
Integrity of the housing
▪
Integrity of the display
▪
Integrity of the sockets
▪
Integrity of the heating plate and connection
button of the water chamber
▪
Legibility of the type plate
▪
Integrity of the mains cable
If the integrity is compromised, the device must not be
used. Contact the technical customer service of
Löwenstein Medical.
Annual testing
Before you carry out the necessary annual tests,
install the ventilation system.
(→
"Start-up" p. 25)
The annual tests may only be carried out by
competent and authorised/certified personnel in
accordance with the maintenance manual of
Löwenstein Medical. Löwenstein Medical accepts no
responsibility for the proper functioning and safety of
the LM 2000 humidifier if the annual tests required by
the maintenance manual have not been carried out
by competent and authorised/certified personnel.
The LM 2000 must undergo the following planned
annual tests every twelve months to maintain its
performance and safety.
Visual inspections
▪
Integrity of the housing
▪
Integrity of the display
▪
Integrity of the sockets
▪
Integrity of the heating plate
▪
Integrity of the attachment symbol of the water
chamber
▪
Legibility of the type plate
▪
Integrity of the mains cable
▪
Integrity of the heating cable