SensoDirect 110

4. Taking measurements

1.  Connect the Salt probe to the probe input socket (Figure 1, 5).
2.  Power on the instrument by pressing the Power ON Button.
3.  Hold the „Probe“ Handle by hand and immerse the Salt Sensor wholly into the meas-

ured solution. Shake the Probe several time to let the air bubble leave away from the 
Salt Sensor until the reading value reach stable. Display will show the salt values (% 
weight) as NaCl.

4.  The Probe Head has build in a temperature sensor for the usage of automatic Temp. 

compensation. If the temperature of measured salt solutions is changed, then it 
should take few minutes to let the display reading reacht the stable value.

5.  Press the „Hold Button“ (Figure 1, 4) will hold the measured value and the LCD will 

indicate a „Hold“ symbol on the display during the measurement. Press the Hold Button 
again to exit the data hold function.

5. Calibration

Figure 2

To calibrate the meter:

1. Prepare a 0.6% weight salt Solution (NaCl)   6000 mg/l (ppm). 

 

For example: 1000g salt Solution contain 6g salt.

2. Hold the Probe Handle by hand and immerse the let the Probe Head wholly into the meas-

ured solution. Shake the Probe several time to let the air bubble leave away from the Probe 
Head until the reading value reach stable. Calibrate the instrument with the calibration 
screws (Figure 2, 6) until display show the value same as 0.60 exactly

3. Alternative Calibrations

Concentration

1000g salt solution 
contains

mg/l (ppm)

Display reading

3%

30 g Salt

30000

3.00

8%

80 g Salt

80000

8.00

rough adjustment

vernier adjustment

Содержание Tintometer SensoDirect Salt110

Страница 1: ...GB DE SensoDirect Salt110 Salt Meter Instruction Manual Lovibond Water Testing Tintometer Group...

Страница 2: ...packen den Lieferumfang Es sollten folgende Bestandteile enthalten sein Lovibond SensoDirect Salt110 Lovibond SensoDirect Salz Elektrode 9 V Batterie Schutzarmierung 7 Schrauben Schraubendreher Kunsts...

Страница 3: ...eratur Automatisch 0 50 C kompensation Betriebstemperatur 0 50 C Umgebungsbedingungen maximal 80 rel Feuchte Stromversorgung 006P DC 9V Batterie Blockbatterie MN1604 PP3 oder gleichwertig Stromverbrau...

Страница 4: ...dung 1 4 dient zum einfrieren des aktuellen Messwertes im Display Es erscheint HOLD im Display Erneutes Dr cken der HOLD Taste kehrt zum Messmodus zur ck 5 Justierung Abbildung 2 Justieren des Ger tes...

Страница 5: ...inen des Zeichens m glich 2 Vor dem Batteriewechsel ist die Schutzarmierung zu entfernen Zum Austausch der Batterie den Batteriefachdeckel an der R ckseite des Ger tes l sen Abbildung 3 3 Batterie ent...

Страница 6: ...n keys Unpacking Please check that the shipment includes the following items Lovibond SensoDirect Salt110 Lovibond SensoDirect Salt Sensor 9V Battery Protective Cover 7 Screws Screwdriver Plastic Cove...

Страница 7: ...atic 0 50 C 32 F 122 F Compensation Operating Temperature 0 50 C 32 F 122 F Operating Humidity Maximum 80 relative humidity Power Supply 006P DC 9V battery heavy duty type MN1604 PP3 or equivalent Pow...

Страница 8: ...Press the Hold Button Figure 1 4 will hold the measured value and the LCD will indicate a Hold symbol on the display during the measurement Press the Hold Button again to exit the data hold function...

Страница 9: ...ow Bat tery indicator first appears 2 Remove the protective cover before replacing the battery To replace the battery remove the Battery Compartment Cover Figure 3 on the rear of the meter 3 Remove th...

Страница 10: ...10 SensoDirect 110...

Страница 11: ...11 SensoDirect 110...

Страница 12: ...Fax 49 0 231 94510 20 sales tintometer de www lovibond com Deutschland The Tintometer Ltd Lovibond House Sun Rise Way Amesbury Salisbury SP4 7GR Tel 44 0 1980 664800 Fax 44 0 1980 625412 sales tintom...

Отзывы: