Troubleshooting
Troubleshooting
Symptom
Action
No 0-20/4-20mA output
Check connection with sensor
Relays off continuously
Check connection with sensor
Check cable at position J3 on low voltage board
Relays on continuously
Check connection with sensor
No green power light
Check fuses
Check AC mains circuit breaker
Check cable at J9 on low voltage board
No USB connection
Check USB power LED1
USB driver not found
Download and install “CP210x Windows Drivers”