34
PCM_V2.1 12.2018
Figure 15:
Modbus RS485/RS232 Connector
Mating Board Connector
The Modbus Wiring Adapter contains 2 slide switches SW1 and SW2 .
SW1 enables the integrated resistor combination . Setting SW1 to “T” side, connects
the terminating resistor across the network . Setting SW1 to “O” side disconnects the
terminating resistor, which can be used to change from RS485 to RS232 .
SW2 is used to select RS485 or RS232 for communication . Setting SW2 to “RS232”
enables R232 . Setting SW2 to “RS485” enables RS485 .
GB
Digital Interface