16
PCM_V2.1 12.2018
3 .
Remove the appropriate hole plugs from the bonding plate .
The two near the back are intended for high voltage use .
Use proper connection hardware to ensure grounding connection to
metal bonding plate . (see Figure 5)
4 .
Connect AC power to the connector labeled J1 (see Figure 6 for
power connector location) . AC power and protective earth wires
must be 12 to 18 AWG . A local disconnect switch or circuit breaker
which meets local electrical code is also required on the AC power
input, and must be located near the PCM .
5 .
Connect Relay outputs to connectors labeled J2, J3, and J4 . Note that
Normally Open (NO) and Normally Closed (NC) contacts are provided .
See Figure 6 for wiring information .
6 .
Reinstall the low voltage board connectors at position J102 and J109
and the red thumb screw .
GB
Installation
High Voltage Installation
DANGER
ELECTRIC SHOCK HAZARD . The barrier must stay in place except when a qualified
installation technician is connecting power or relay outputs .
CAUTION
ENSURE PROPER GROUNDING . Connect metal bonding plate to protective earth
for proper operation .
WARNING
POTENTIAL FIRE HAZARD . The relay contacts are not fused and are rated at 5A max .
External devices should have a current limiting device to keep the current below 5A .
Figure 5:
Grounding connection
A
Содержание Tintometer PTV Series
Страница 2: ...THIS PAGE INTENTIONALLY BLANK ...
Страница 10: ...THIS PAGE INTENTIONALLY BLANK ...
Страница 19: ...19 PCM_V2 1 12 2018 GB Installation J6 J5 Figure 8 4 20 mA connections ...
Страница 21: ...21 PCM_V2 1 12 2018 Figure 9 12 pin connection GB Installation ...
Страница 30: ...THIS PAGE INTENTIONALLY BLANK ...
Страница 38: ...THIS PAGE INTENTIONALLY BLANK ...
Страница 39: ...THIS PAGE INTENTIONALLY BLANK ...