8
PCM_V1 07.2017
DANGER
DISCONNECT POWER BEFORE SERVICING! Only a qualified installation technician
should connect AC power to the PCM
1 .
Remove the red thumb screw, and connectors at position J3 and J9
2 .
Remove the low voltage circuit board and high voltage barrier
GB
Installation
High Voltage Installation
NOTICE
A local disconnect mounted in a suitable and easily reached location must be included
in the installation of this equipment . The local disconnect must be marked as the
disconnecting device for this equipment .
NOTICE
This unit was tested with metallic conduit as per UL 61010-1 with a bending moment of
34Nm (300 lb-in) at a horizontal distance of 300mm (12") . Do not exceed this bending
moment