background image

SD %=

pH

SD 50

pH

7. Delete Stored Data

Recall Data (see 6.)
Press and hold (3 sec.)

CLEAR will flash

Confirm deletion of stored data
This erases all stored data

Automatic return to normal display 

8.  Select Buffer System

Power ON

Press and hold (3 sec.)

Press 1 x

Current pH system will flash
Press to confirm

Toggle to change / select (7.00 

or 6.86)
Save and exit

9. Set Date/Time display  
 type

Power ON

Press and hold (3 sec.)

Press 4 x

Date is displayed

Confirm

Toggle to 12 hour clock (am/pm) or 

24 hour clock

Save and exit

4. Preparation of buffer  
 solution

1. Fill a clean beaker with 20 ml distilled water.
2. Add one buffer tablet of the required value 

and shake or crush with a clean stirring rod 

until the tablet dissolves.
3. Allow to stand for about two minutes to 

ensure that the buffer is completely dissolved 

and to allow any insoluble particles to settle.
4. Immerse the electrode into the buf-

fer solution and carry out calibration  

(see 3. Calibration)
5. Remove the electrode and rinse it tho-

roughly
6. Use the buffer solution only once

Accuracy

The maximum tolerance of fresh buffer solu-

tions is ±0.05 pH.

5. Selection of pH and mV  
 mode

Toggle to select mode

Note: mode mV does NOT show 

ORP on the display, only millivolt.

6. Recall Data

Power ON

Press and hold (3 sec.)

Most recent memory setting will 

flash
Press to recall

Most recent memory setting is 

displayed

Toggle to other stored values

Press to exit

Press and hold (3 sec.) to start 

calibration

CAL 1 will flash

Wait until     disappears

Press 1 sec.
Calibration will take place

CAL 2 will flash

Exit for 1 point calibration or go to 

2nd point calibration

Rinse electrode with distilled 

water and immerse into pH 4  

L

 buffer solution.

CAL 2 will flash

Wait until    disappears

Press and hold (1 sec.)

Calibration will take place
CAL 3 will flash

Exit for 2 point calibration or go to 

3rd point calibration

Rinse electrode with distilled 

water and immerse into pH 10  

H

 buffer solution

CAL 3 will flash

Wait until   disappears

Press and hold (1 sec.)

Calibration will take place

3 point calibration is finished

L

M

H

 will appear on screen and 

show type of calibration (1 point to 

3 point calibration)

1. Power ON/OFF

Press briefly

If ! is displayed,              
press to deactivate

Automatically turns off after  

8 minutes of inactivity

Press and hold (3 sec.) to turn the 

instrument off

Press to activate / de-activate  

backlight

2. Taking a reading

Remove protective cap and immerse 

the instrument into the sample to be 

measured so that the instrument is 

submersed to the sealing ring level.

Reading is displayed

Press briefly ! to freeze / unfreeze

Press and hold (3 sec.) to store data
While pressing button, memory 

listing (e.g. 01) appears

3. Calibration

Rinse electrode thoroughly with 

distilled water before using in each 

buffer solution. 
Re-calibrate according to individual 

recommendations.
Use Lovibond

®

 Standard Solutions or 

pH buffer tablets in accordance with 

the kit manual (see 4.)

Remove protective cap from the 

probe. Turn ON, rinse electrode 

with distilled water and immerse 

the probe into pH 7 

M

 solution.