background image

SD 50

pH

Lock/Unlock Data!

C. Select pH/mV

D. Backlight ON/OFF

____________________________________________________________________________________________

 

Note: Icons/Characters in 

RED

 meaning "

BLINK

          

: Hit once  /     

: Press and hold  /     

: Release 

F. Store Data 

1. In lock mode: 

 

 

2s 

 

 

 

2. In measurement mode: 

 

 

2s 

 

 

 (or 

 

 

2s 

 

 

 ) 

3. When the storage is full, it will overwrite the oldest data. 

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- 

 
G. System Setup 

Menu selection: 

 

 

2s 

 

 

 (latest memory#) 

 

 

 (

7.00

 or 

6.86

 

 

 (

dd-mm-yy/24hr

 or 

mm-dd-yy/12hr

 

 

 (

°C

 or 

°F

 

 

 (

31-12  24

 or 

12-31  12ampm

 

 

..... 

 

To enter: 

 

 

!

F. pH Buffer  7.00 -> 4.00 -> 10.00 

MEASUREMENT

pH CALIBRATION

Immerse the probe 
into pH7 buffer 
solution.

Exit here for one 
point calibration

Exit here for two 
point calibration

or go for second point calibration as follows:

or go for third point calibration as follows:  

wait

wait

wait

Clean the probe first, 
then immerse into pH4 
buffer solution.

Clean the probe first, 
then immerse into 
pH10 buffer solution.

(4)          (7)          (10)

on the Display

Remove protective cap from the probe

E. Store Data

Three point calibration 
finished.

L       M         H 

Low, Middle, High

Press

Press for 

3 seconds

Insert the batteries

unlock

off

off

pH

unlock

off

Automatically turns off after 8 minutes of 
inactivity.

Possible slight delay after pressing 
before freeze of the data (!).

Mode „mV“ does not show ORP on the display, 
only milivolts.

to switch, unfreeze data (see B).

After 3 sec. the file no. temporarily appears 
on the display, e.g. 02. 
Once the storage has reached max. capacity, 
the data will overwrite the oldest files.

off

pH

lock

on

on

mV

• Re-Calibrate according to individual recommendations! Rinse the electrode with destilled water after each measurement.
• Choose Lovibond Standard Solutions or produce them with buffer tablets according to the kit manual (see backside).

Remove protective cap from the probe.
Power ON (see A)

GB

Characters in 

RED

 represent „

Flashing symbols

“.

4

5

6

+

+

+

1

2

3

+

-

-

-

-