ENGLISH
• Calculation
Combined Chlorine
= Total Chlorine - Free Chlorine
Technical Data
Free chlorine: 0.05 - 6.0 mg/l Cl
2
Total chlorine: 0.05 - 6.0 mg/l Cl
2
(Accuracy Chlor: 0-1 mg/l ± 0.2; 1-2 mg/l ±0.3; 2-3 mg/l ±0.4; 3-6 mg/l ±0.5)
Bromine:
0.1 - 13.5 mg/l Br
(Accuracy:
0-2 mg/l ± 0.5; 2-4 mg/l ±0.7; 4-7 mg/l ±0.9; 7-13 mg/l ±1.2mg/l)
pH value:
6.5 - 8.4 pH
(Accuracy ±0.2 pH)
Cyanuric Acid: 1 - 80 mg/l Cys
(Accuracy ±10 mg/l)
Alkalinity:
10 - 500 mg/l CaCO
3
(Accuracy ±50 mg/l)
Light source:
LED, filter (
λ
= 528 nm)
Battery:
9 V-block battery (Life 600 tests).
Auto-OFF:
Automatic switch off 5 minutes after last keypress
Ambient
temperature: 0 – 40°C, relative humidity: 30 – 90%,
conditions:
non-condensing
CE:
DIN EN 55 022, 61 000-4-2, 61 000-4-8,
50 082-2, 50 081-1, DIN V ENV 50 140, 50 204
• Maintainance
1. Keep the instrument in a clean condition.
2. No solvent or abrasive materials should be used to clean the instrument.
3. Let the instrument dry when not used, store under ambient
environmental conditions.
4. Remove battery if instrument is longer unused.
• Trouble shooting
1. Use only tablets in white foil.
2. Touching the tablets can cause contamination.
3. Close the grey lid on top of the sample chamber before
zeroing/reading.
4. Always ensure that the sample and the lid are thoroughly rinsed
when changing from one test to the other.
• Fitting the battery (9 volts, 6LR61)
BAT
Change the batteries with this display indication. Check
position of the black O-Ring when opening/ closing the battery
chamber lid!
6
7
9
TEST
water
tight
position