51
MaxiDirect_2c 03/2010
1.1 Methods
12. Place the vial in the sample chamber making sure that
the marks are aligned.
13. Press
TEST
key.
Wait for a
reaction period of 3 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in:
mg/l free Chlorine
mg/l combined Chlorine
mg/l total Chlorine
Notes:
See page 41
*,** mg/l free Cl
*,** mg/l comb. Cl
*,** mg/l total Cl
T1 accepted
prepare T2
press TEST
Countdown
3:00