259
MaxiDirect_2c 03/2010
3.4 Technical data
Display
Graphic Display
Serial Interface
IR interface for data transfer
RJ45 connector for internet updates (see chapter 2.5.3)
Light source
light-emitting diode – photosensor – pair arrangement
in a transparent measurement chamber
Wavelength ranges:
1 = 530 nm IF
= 5 nm
2 = 560 nm IF
= 5 nm
3 = 610 nm IF
= 6 nm
4 = 430 nm IF
= 5 nm
5 = 580 nm IF
= 5 nm
6 = 660 nm IF
= 5 nm
IF = Interference filter
Wavelength accuracy
± 1 nm
Photometric accuracy*
2% FS (T = 20°C – 25°C)
Photometric resolution
0.005 A
Protection
conforming to IP 68 (1 h, 0.1 m)
Operation
Acid and solvent resistant touch-sensitive keyboard with
integral beeper as acoustic indicator.
Power supply
4 batteries (Type AA/LR 6);
lifetime: approx. 26 hours continuous use or 3500 tests
Auto off
20 minutes after last function,
30 seconds acoustical signal before switch off
Dimensions
approx. 210 x 95 x 45 mm (unit)
approx. 395 x 295 x 106 mm (case)
Weight (unit)
approx. 450 g
Working condition
5 – 40°C at max. 30–90% relative humidity
(without condensation)
Language options
English, German, French, Spanish, Italian, Portuguese,
Polish; further languages via Internet Update
Storage capaity
ca. 1000 data sets
* measured with standard solutions
Subject to technical modification!
To ensure maximum accuracy of test results, always use the reagent systems
supplied by the instrument manufacturer.
Содержание Maxi Direct
Страница 1: ...Photometer System axi GB Instruction manual ...
Страница 2: ......
Страница 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...
Страница 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 125: ...123 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 Some solids may not dissolve 2 Conversion mg l NO3 mg l N x 4 43 3 N NO3 ...
Страница 143: ...141 MaxiDirect_2c 03 2010 ...
Страница 187: ...185 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 If Sulfate is present a cloudy solution will appear ...
Страница 189: ...187 MaxiDirect_2c 03 2010 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will appear ...
Страница 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 206: ...204 MaxiDirect_2c 03 2010 ...
Страница 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...
Страница 210: ...208 MaxiDirect_2c 03 2010 ...
Страница 258: ...256 MaxiDirect_2c 03 2010 ...
Страница 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...
Страница 267: ...265 MaxiDirect_2c 03 2010 ...