24
MaxiDirect_2c 03/2010
1.1 Methods
Ammonia
with Vario Powder Pack
0.01 – 0.8 mg/l N
Use two clean vials (24 mm Ø) and mark one as blank for
zeroing.
1. Fill a clean vial (24 mm Ø) with
10 ml of deionised
water
(this is the blank).
2. Fill the other clean vial (24 mm Ø) with
10 ml of the
water sample
(this is the sample).
3. Add the contents of
one Vario Ammonia Salicylate
F10 Powder Pack
straight from the foil to each vial.
4. Close the vials with the caps and shake to mix the
contents.
5. Press
[ ]
key.
Wait for a
reaction period of 3 minutes
.
After the reaction period is finished proceed as follows:
6. Add the contents of
one Vario Ammonia Cyanurate
F10 Powder Pack
straight from the foil to each sam-
ple.
7. Close the vials tightly with the caps and shake to mix
the contents.
8. Press
[ ]
key.
Wait for a
reaction period of 15 minutes
.
After the reaction period is finished proceed as follows:
9. Place the vial (the blank) in the sample chamber making
sure that the marks are aligned.
10. Press
ZERO
key.
11. Remove the vial from the sample chamber.
12. Place the vial (the sample) in the sample chamber
making sure that the marks are aligned.
13. Press
TEST
key.
The result is shown in the display in mg/l Ammonia as N.
1
2
3
4
Countdown 1
3:00
start:
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown 2
15:00
start:
6
2
Содержание Maxi Direct
Страница 1: ...Photometer System axi GB Instruction manual ...
Страница 2: ......
Страница 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...
Страница 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 125: ...123 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 Some solids may not dissolve 2 Conversion mg l NO3 mg l N x 4 43 3 N NO3 ...
Страница 143: ...141 MaxiDirect_2c 03 2010 ...
Страница 187: ...185 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 If Sulfate is present a cloudy solution will appear ...
Страница 189: ...187 MaxiDirect_2c 03 2010 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will appear ...
Страница 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 206: ...204 MaxiDirect_2c 03 2010 ...
Страница 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...
Страница 210: ...208 MaxiDirect_2c 03 2010 ...
Страница 258: ...256 MaxiDirect_2c 03 2010 ...
Страница 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...
Страница 267: ...265 MaxiDirect_2c 03 2010 ...