251
MaxiDirect_2c 03/2010
Initialise User Method System (Polynomials & Concentration)
Power loss will cause incoherent data. The user method system must be initialised with this
mode function to set it to a predefined state.
ATTENTION:
All stored user methods (polynomial & concentration) are deleted with initialisation.
Press [MODE], [Shift] + [6][9] keys.
Confirm with [ ] key.
The display shows:
Confirm with [ ] key.
The query is displayed:
• Press [Shift] + [1] keys to start initialisation.
• Press [Shift] + [0] keys to to cancel without initialisation.
The instrument goes back to mode menu automatically.
<User m. init>
Start:
Initialising?
YES: 1, NO: 0
0
Mode
9
6
1
Содержание Maxi Direct
Страница 1: ...Photometer System axi GB Instruction manual ...
Страница 2: ......
Страница 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...
Страница 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 125: ...123 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 Some solids may not dissolve 2 Conversion mg l NO3 mg l N x 4 43 3 N NO3 ...
Страница 143: ...141 MaxiDirect_2c 03 2010 ...
Страница 187: ...185 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 If Sulfate is present a cloudy solution will appear ...
Страница 189: ...187 MaxiDirect_2c 03 2010 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will appear ...
Страница 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 206: ...204 MaxiDirect_2c 03 2010 ...
Страница 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...
Страница 210: ...208 MaxiDirect_2c 03 2010 ...
Страница 258: ...256 MaxiDirect_2c 03 2010 ...
Страница 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...
Страница 267: ...265 MaxiDirect_2c 03 2010 ...