244
MaxiDirect_2c 03/2010
User Concentration Methods
It is possible to enter and store up to 10 User Concentration Methods.
Therefor you need 2 to 14 standards of known concentration and one blank (deionised water
or reagent blank value). The Standards should be measured with increasing concentrations
and from the brightest to the darkest colouration.
The measuring range for „Underrange“ and „Overrange“ is defined with –2600 mAbs* and
+2600 mAbs*. After selection of a method the concentration of the lowest and highest used
standard is displayed as measuring range. The operation range should be within this range
to achieve best results.
*1000 mAbs = 1 Abs = 1 E (displayed)
Entering a User Concentration:
Press [MODE], [Shift] + [6][4] keys.
Confirm with [ ] key.
Entry Procedure:
The display shows:
Enter a method number in the range from 850 to 859,
e.g.: [Shift] + [8][5][0]
Confirm with [ ] key.
Note:
if the entered number has already been used to save a
concentration the display shows the query:
• Press [Shift] + [0] or [ESC] keys to go back to method no.
query.
• Press [Shift] + [1] keys to start entry mode.
Enter the required wavelength, e.g.: [Shift] + [2] for
560 nm.
Press [ ] or [ ] keys to select the required unit.
Confirm with [ ] key.
< User concentr.>
choose no.: ____
(850-859)
choose unit:
>>
mg/l
g/l
mmol/l
mAbs
µg/l
E
A
%
Overwrite conc. meth.?
YES: 1, NO: 0
wavelength:
1: 530 nm
4: 430 nm
2: 560 nm
5: 580 nm
3: 610 nm
6: 660 nm
Mode
4
6
2
8
0
5
Содержание Maxi Direct
Страница 1: ...Photometer System axi GB Instruction manual ...
Страница 2: ......
Страница 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...
Страница 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 125: ...123 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 Some solids may not dissolve 2 Conversion mg l NO3 mg l N x 4 43 3 N NO3 ...
Страница 143: ...141 MaxiDirect_2c 03 2010 ...
Страница 187: ...185 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 If Sulfate is present a cloudy solution will appear ...
Страница 189: ...187 MaxiDirect_2c 03 2010 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will appear ...
Страница 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 206: ...204 MaxiDirect_2c 03 2010 ...
Страница 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...
Страница 210: ...208 MaxiDirect_2c 03 2010 ...
Страница 258: ...256 MaxiDirect_2c 03 2010 ...
Страница 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...
Страница 267: ...265 MaxiDirect_2c 03 2010 ...