201
MaxiDirect_2c 03/2010
1.2.2 Cleaning of vials and accessories for analysis
Vials, caps and stirring rods should be cleaned thoroughly
after each analysis
to prevent
interferences.
Procedure:
Clean vials and accessories after each analysis as soon as possible.
a. Clean vials and accessories with laboratory detergent
(e.g. Extran
®
MA 02 (neutral, phosphatic), Extran
®
MA 03 (alkaline, phosphate-free)
from Merck KGaA).
b. Rinse thoroughly with tap water.
c. On demand (see Notes) perform special cleaning as required, e.g.: rinse with diluted
Hydrochloric acid solution.
d. Rinse thoroughly with deionised water.
1.2.3 Guidelines for photometric measurements
1. Vials, caps and stirring rods should be cleaned thoroughly after each analysis to prevent
interferences. Even minor reagent residues can cause errors in the test result.
2. The outside of the vial must be clean and dry before starting the analysis. Clean the
outside of the vials with a towel. Fingerprints or other marks will be removed.
3. If there is no defined vial for the blank, the zeroing and the test must be carried out with
the same vial as there may be slight differences in optical performance between vials.
4. The vials must be positioned in the sample chamber for zeroing and test with the mark
on the vial aligned with the mark on the instrument.
Correct position of the vial (Ø 24 mm):
Содержание Maxi Direct
Страница 1: ...Photometer System axi GB Instruction manual ...
Страница 2: ......
Страница 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...
Страница 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 125: ...123 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 Some solids may not dissolve 2 Conversion mg l NO3 mg l N x 4 43 3 N NO3 ...
Страница 143: ...141 MaxiDirect_2c 03 2010 ...
Страница 187: ...185 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 If Sulfate is present a cloudy solution will appear ...
Страница 189: ...187 MaxiDirect_2c 03 2010 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will appear ...
Страница 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 206: ...204 MaxiDirect_2c 03 2010 ...
Страница 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...
Страница 210: ...208 MaxiDirect_2c 03 2010 ...
Страница 258: ...256 MaxiDirect_2c 03 2010 ...
Страница 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...
Страница 267: ...265 MaxiDirect_2c 03 2010 ...