179
MaxiDirect_2c 03/2010
1.1 Methods
11. Press
ZERO
key (blank is already placed in the sample
chamber – see point 8).
Wait for a
reaction period of 2 minutes
.
After the reaction period is finished the zero-reading
starts automatically.
12. Remove the vial from the sample chamber.
13. Place the vial (the sample) in the sample chamber ma-
king sure that the marks are aligned.
14. Press
TEST
key.
The result is shown in the display in mg/l Silica.
Notes:
1. Close the vials with the cap immediately after adding the Vario Molybdate 3 reagent
solution, otherwise low readings may result.
2. The given reaction time of 4 minutes refers to a water sample temperature of 20°C.
At 30°C a reaction time of 2 minutes, at 10°C a reaction time of 8 minutes are required.
3. The given reaction time of 1 minute refers to a water sample temperature of 20°C.
At 30°C a reaction time of 30 seconds, at 10°C a reaction time of 2 minutes are required.
4. Interferences:
Substance
Interference
Iron
large amounts interfere
Phosphate
does not interfere at concentrations less than 50 mg/l PO
4
at 60 mg/l PO
4
the interference is approx. – 2%
at 75 mg/l PO
4
the interference is approx. – 11%
Sulfide
interferes at all levels
Occasionally water samples contain forms of silica which reacts very slowly with
Molybdate. The nature of these forms is not known.
A pre-treatment with Sodium hydrogencarbonate and then with Sulfuric Acid will
make these forms reactive to Molybdate (pre-treatment is given in “Standard Methods
for the Examination of Water and Wastewater” under “ Silica Digestion with Sodium
Bicarbonate”).
5. SiO
2
Si
Countdown
2:00
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Содержание Maxi Direct
Страница 1: ...Photometer System axi GB Instruction manual ...
Страница 2: ......
Страница 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...
Страница 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 125: ...123 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 Some solids may not dissolve 2 Conversion mg l NO3 mg l N x 4 43 3 N NO3 ...
Страница 143: ...141 MaxiDirect_2c 03 2010 ...
Страница 187: ...185 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 If Sulfate is present a cloudy solution will appear ...
Страница 189: ...187 MaxiDirect_2c 03 2010 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will appear ...
Страница 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 206: ...204 MaxiDirect_2c 03 2010 ...
Страница 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...
Страница 210: ...208 MaxiDirect_2c 03 2010 ...
Страница 258: ...256 MaxiDirect_2c 03 2010 ...
Страница 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...
Страница 267: ...265 MaxiDirect_2c 03 2010 ...