176
MaxiDirect_2c 03/2010
1.1 Methods
Silica/Silicon dioxide
with Tablet
0.05 – 4 mg/l SiO
2
1. Fill a clean vial (24 mm Ø) with
10 ml of the water sample
,
close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add
one SILICA No. 1 tablet
straight from the foil
to the water sample and crush the tablet using a clean
stirring rod.
6. Close the vial tightly with the cap and swirl several times
until the tablet is dissolved.
7. Press [ ] key.
Wait for a
reaction period of 5 minutes
.
After the reaction period is finished proceed as follows:
8. Add
one SILICA PR tablet
straight from the foil to the
same water sample and crush the tablet using a clean
stirring rod.
9. Add
one SILICA No. 2 tablet
straight from the foil to
the same water sample and crush the tablet using a
clean stirring rod.
10. Close the cap tightly and swirl several times until the
tablets are dissolved.
11. Place the vial in the sample chamber making sure that
the marks are aligned.
12. Press
TEST
key.
Wait for a
reaction period of 1 minute
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Silica.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
1:00
Countdown
5:00
start:
5
0
3
Содержание Maxi Direct
Страница 1: ...Photometer System axi GB Instruction manual ...
Страница 2: ......
Страница 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...
Страница 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 125: ...123 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 Some solids may not dissolve 2 Conversion mg l NO3 mg l N x 4 43 3 N NO3 ...
Страница 143: ...141 MaxiDirect_2c 03 2010 ...
Страница 187: ...185 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 If Sulfate is present a cloudy solution will appear ...
Страница 189: ...187 MaxiDirect_2c 03 2010 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will appear ...
Страница 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 206: ...204 MaxiDirect_2c 03 2010 ...
Страница 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...
Страница 210: ...208 MaxiDirect_2c 03 2010 ...
Страница 258: ...256 MaxiDirect_2c 03 2010 ...
Страница 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...