162
MaxiDirect_2c 03/2010
1.1 Methods
Phosphonates
Persulfate UV oxidation method
with Vario Powder Pack
0 – 125 mg/l (see Table 1)
1. Choose the appropriate sample volume from table 1
(see following pages).
2. Pipette the chosen sample volume into a clean 50 ml
graduated cylinder. If necessary fill up with deionised
water to the 50 ml mark and mix well.
3. Fill a clean vial (24 mm Ø) with
10 ml of the prepared
water sample
(this is the blank).
4. Transfer
25 ml of the prepared water sample
into
the digestion vial.
5. Add the contents of
one Vario Potassium Persulfate
F10 Powder Pack
straight from the foil to the digestion
vial.
6. Close the digestion vial tightly with the cap and swirl
until the reagent is dissolved completely.
7. Insert the UV lamp into the digestion vial (Note 3, 4, 5).
CAUTION: Wear UV safety goggles!
8. Switch the UV lamp on and wait for a
reaction period
of 10 minutes
.
9. After the reaction period is finished switch the UV lamp
off and remove the lamp from the vial.
10. Fill a second vial (24 mm Ø) with
10 ml of the digested
sample
(this is the sample).
11. Add the contents of
one Vario Phosphate Rgt. F10
Powder Pack
straight from the foil into each vial (blank
and sample).
12. Close the vials tightly with the cap and swirl gently
several times (30 sec.).
(Note 6)
3
1
6
Countdown 1
10:00
start:
1
2
3
4
Содержание Maxi Direct
Страница 1: ...Photometer System axi GB Instruction manual ...
Страница 2: ......
Страница 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...
Страница 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 125: ...123 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 Some solids may not dissolve 2 Conversion mg l NO3 mg l N x 4 43 3 N NO3 ...
Страница 143: ...141 MaxiDirect_2c 03 2010 ...
Страница 187: ...185 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 If Sulfate is present a cloudy solution will appear ...
Страница 189: ...187 MaxiDirect_2c 03 2010 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will appear ...
Страница 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 206: ...204 MaxiDirect_2c 03 2010 ...
Страница 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...
Страница 210: ...208 MaxiDirect_2c 03 2010 ...
Страница 258: ...256 MaxiDirect_2c 03 2010 ...
Страница 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...
Страница 267: ...265 MaxiDirect_2c 03 2010 ...