1.1 Methods

Acid demand to pH 4.3 

with Tablet

0.1 – 4 mmol/l

1.  Fill a clean vial (24 mm Ø) with 

10 ml of the water  

 

sample

, close tightly with the cap.

2.  Place the vial in the sample chamber making sure that 
 

the   marks are aligned.

3.  Press 

ZERO

 key.

4.  Remove the vial from the sample chamber.

5.  Add 

one  ALKA-M-PHOTOMETER  tablet

  straight  

 

from the foil to the water sample and crush the tablet  

 

using a clean stirring rod.

6.  Close the vial tightly with the cap and swirl several times  
 

until the tablet is dissolved.

7.  Place the vial in the sample chamber making sure that  
 

the   marks are aligned.

8.  Press 

TEST

 key.

 

The result is shown in the display as Acid demand to 

 

pH 4.3 in mmol/l.

 prepare Zero
 press ZERO

 Zero accepted
 prepare Test
 press TEST 

2

0

Содержание Maxi Direct

Страница 1: ...Photometer System axi GB Instruction manual ...

Страница 2: ......

Страница 3: ...ruction manual before unpacking setting up or using the photometer Please read the method description completely before performing the test Be aware of the risks of using the required reagents by reading the MSDS Material Safety Data Sheets Failure could result in serious injury to the operator or damage to the instrument MSDS www tintometer de CAUTION The accuracy of the instrument is only valid ...

Страница 4: ...ne 40 Chlorine with tablet differentiated determination free combined total 42 free Chlorine 44 total Chlorine 45 Chlorine with liquid reagent differentiated determination free combined total 46 free Chlorine 48 total Chlorine 49 Chlorine powder pack differentiated determination free combined total 50 free Chlorine 52 total Chlorine 53 Chlorine dioxide 54 in presence of Chlorine 56 in absence of C...

Страница 5: ...ide 100 Iodine 102 Iron 104 Iron with tablet 106 Iron powder pack 108 Iron TPTZ powder pack 110 Manganese with tablet 112 Manganese LR powder pack 114 Manganese HR powder pack 116 Molybdate with tablet 118 Molybdate HR powder pack 120 Nitrate 122 Nitrite with tablet 124 Nitrite LR powder pack 126 Nitrogen total LR 128 Nitrogen total HR 130 Oxygen active 132 Oxygen dissolved 134 Ozone 136 in presen...

Страница 6: ...178 Silica HR PP 180 Sodium hypochlorite 182 Sulfate with tablet 184 Sulfate powder pack 186 Sulfide 188 Sulfite 190 Suspended Solids 192 Turbidity 194 Urea 196 Zinc 198 1 2 Important notes 200 1 2 1 Correct use of reagents 200 1 2 2 Cleaning of vials and accessories for analysis 201 1 2 3 Guidelines for photometric measurements 201 1 2 4 Sample dilution techniques 203 1 2 5 Correcting for volume ...

Страница 7: ... 3 9 Printing results Infra Red Interface Module optional 215 2 3 10 Perform additional measurements 215 2 3 11 Selecting a new method 216 2 3 12 Measure absorbance 216 2 4 Photometer settings Table of Mode Functions 217 2 4 1 blank because of technical requirements 2 4 2 Instrument basic settings 1 218 2 4 3 Printing of stored results 222 2 4 4 Recall delete stored results 227 2 4 5 Calibration 2...

Страница 8: ...ng 258 3 2 Delivery contents 258 3 3 blank because of technical requirements 3 4 Technical data 259 3 5 Abbreviations 260 3 6 Troubleshooting 261 3 6 1 Operating messages in the display error display 261 3 6 2 General 263 3 7 Declaration of CE Conformity 264 ...

Страница 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...

Страница 10: ...63 Chloramine mono PP PP liquid 0 04 4 50 mg l Cl2 Indophenol 660 34 90 Chloride T tablet 0 5 25 mg l Cl Silver nitrate turbidity 530 38 100 Chlorine T tablet 0 01 6 mg l Cl2 DPD 1 2 3 530 40 42 101 Chlorine L liquid 0 02 4 mg l Cl2 DPD 1 2 3 530 40 46 110 Chlorine PP PP 0 02 2 mg l Cl2 DPD 1 2 530 40 50 120 Chlorine dioxide T tablet 0 05 11 mg l ClO2 DPD Glycine 1 2 530 54 105 Chlorine HR Kl T ta...

Страница 11: ... l Fe 1 10 Phenan troline 3 530 104 108 223 Iron TPTZ PP PP 0 02 1 8 mg l Fe TPTZ 580 104 110 240 Manganese T tablet 0 2 4 mg l Mn Formaldoxime 530 112 242 Manganese LR PP PP liquid 0 01 0 7 mg l Mn PAN 560 114 243 Manganese HR PP PP liquid 0 1 18 mg l Mn Periodate oxidation 2 530 116 250 Molybdate T tablet 1 50 mg l MoO4 Thioglycolate 4 430 118 252 Molybdate HR PP PP 0 5 66 mg l MoO4 Mercaptoacet...

Страница 12: ...rsulfate UV Oxidation 660 162 329 pH Value LR T tablet 5 2 6 8 Bromocresolpurple5 560 166 330 pH Value T tablet 6 5 8 4 Phenolred5 560 168 331 pH Value L liquid 6 5 8 4 Phenolred5 560 170 332 pH Value HR T tablet 8 0 9 6 Thymolblue 5 560 172 340 Potassium T tablet 0 7 12 mg l K Tetraphenylborate Turbidity4 430 174 350 Silica T tablet 0 05 4 mg l SiO2 Silicomolybdate 2 3 660 176 351 Silica LR PP PP...

Страница 13: ...e the amount present in the sample itself These supplementary results make it possible to estimate the actual concentration of the original sample by comparison Literature The reagent formulations are based on internationally recognised test methods Some are described in national and or international guidelines Deutsche Einheitsverfahren zur Wasser Abwasser und Schlammuntersuchung 1 Standard Metho...

Страница 14: ... vial from the sample chamber 5 Add one ALKA M PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display as Acid demand to pH 4...

Страница 15: ...xiDirect_2c 03 2010 1 1 Methods Notes 1 The terms total Alkalinity Alkalinity m m Value and Acid demand to pH 4 3 are identical 2 For accurate results exactly 10 ml of water sample must be taken for the test ...

Страница 16: ...4 Remove the vial from the sample chamber 5 Add one ALKA M PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display as total A...

Страница 17: ... results exactly 10 ml of water sample must be taken for the test 3 Conversion table Acid demand to pH 4 3 DIN 38 409 KS4 3 German dH English eH French fH 1 mg l CaCO3 0 02 0 056 0 07 0 1 Carbonate hardness reference Hydrogencarbonate anions Example 10 mg l CaCO3 10 mg l x 0 056 0 56 dH 10 mg l CaCO3 10 mg l x 0 02 0 2 mmol l 4 CaCO3 dH eH fH aH ...

Страница 18: ... Remove the vial from the sample chamber 5 Add one ALKA P PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display as Alkalini...

Страница 19: ...carbonate m Carbonate 0 Hydroxide 0 Case 2 Alkalinity p 0 and Alkalinity m 2p Hydrogen carbonate m 2p Carbonate 2p Hydroxide 0 Case 3 Alkalinity p 0 and Alkalinity m 2p Hydrogen carbonate 0 Carbonate 2m 2p Hydroxide 2p m 1 1 Methods Notes 1 The terms Alkalinity p p Value and Alkalinity to pH 8 2 are identical 2 For accurate test results exactly 10 ml of water sample must be taken for the test 3 Th...

Страница 20: ... clean stirring rod dissolve the tablet 6 Add one ALUMINIUM No 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl gently several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 5 minu...

Страница 21: ...gnificant unless the water has fluoride added artificially In this case the following table should be used Fluoride mg l F Displayed value Aluminium mg l Al 0 05 0 10 0 15 0 20 0 25 0 30 0 2 0 05 0 11 0 16 0 21 0 27 0 32 0 4 0 06 0 11 0 17 0 23 0 28 0 34 0 6 0 06 0 12 0 18 0 24 0 30 0 37 0 8 0 06 0 13 0 20 0 26 0 32 0 40 1 0 0 07 0 13 0 21 0 28 0 36 0 45 1 5 0 09 0 20 0 29 0 37 0 48 Example If the...

Страница 22: ...od is finished proceed as follows 5 Add the contents of one Vario Hexamine F20 Powder Pack straight from the foil to the same water sample 6 Dissolve the powder using a clean stirring rod 7 Add 1 drop of Vario Aluminum ECR Masking Reagent in the vial marked as blank 8 Add 10 ml of the prepared water sample to the vial this is the blank 9 Add the remaining 10 ml of the prepared water sample in the ...

Страница 23: ...ature must be between 20 C and 25 C 3 A low test result may be given in the presence of Fluorides and Polyphosphates The effect of this is generally insignificant unless the water has fluoride added artificially In this case the following table should be used Fluoride mg l F Displayed value Aluminium mg l Al 0 05 0 10 0 15 0 20 0 25 0 30 0 2 0 05 0 11 0 16 0 21 0 27 0 32 0 4 0 06 0 11 0 17 0 23 0 ...

Страница 24: ...ng a clean stirring rod 6 Add one AMMONIA No 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 10 minutes After the reaction p...

Страница 25: ...development At a temperature below 20 C the reaction period is 15 minutes 4 Sea water samples Ammonia conditioning reagent is required when testing sea water or brackish water samples to prevent precipitation of salts Fill the test tube with the sample to the 10 ml mark and add one level spoonful of Conditioning Powder Mix to dissolve then continue as described in the test instructions 5 Conversio...

Страница 26: ...d is finished proceed as follows 6 Add the contents of one Vario Ammonia Cyanurate F10 Powder Pack straight from the foil to each sam ple 7 Close the vials tightly with the caps and shake to mix the contents 8 Press key Wait for a reaction period of 15 minutes After the reaction period is finished proceed as follows 9 Place the vial the blank in the sample chamber making sure that the marks are al...

Страница 27: ...ng a total Iron test b add the same iron concentration as determined to the deionised water step 1 The interference will be blanked out successfully Magnesium greater than 6000 mg l CaCO3 Nitrate greater than 100 mg l NO3 N Nitrite greater than 12 mg l NO2 N Phosphate greater than 100 mg l PO4 P Sulfate greater than 300 mg l SO4 Sulfide intensifies the colour Glycine Hydrazine Colour Turbidity Les...

Страница 28: ...er Pack straight from the foil into each vial 5 Close the vials tightly with the caps and swirl several times to dissolve the powder 6 Press key Wait for a reaction period of 20 minutes After the reaction period is finished proceed as follows 7 Place the vial the blank in the sample chamber making sure that the marks are l aligned 8 Press ZERO key 9 Remove the vial from the sample chamber 10 Place...

Страница 29: ...one drop of 0 1 mol l Sodium thiosulfate for each 0 3 mg l Cl2 in a one litre water sample 3 Iron interferes with the test The interferences will be eliminated as follows Determine the amount of total iron present in the water sample To produce the blank add an iron standard solution with the same iron concentration to the vial point 1 instead of deionised water 4 Conversion mg l NH4 mg l N x 1 29...

Страница 30: ...er Pack straight from the foil into each vial 5 Close the vials tightly with the caps and swirl several times to dissolve the powder 6 Press key Wait for a reaction period of 20 minutes After the reaction period is finished proceed as follows 7 Place the vial the blank in the sample chamber making sure that the marks are l aligned 8 Press ZERO key 9 Remove the vial from the sample chamber 10 Place...

Страница 31: ...dd one drop of 0 1 mol l Sodium thiosulfate for each 0 3 mg l Cl2 in a one litre water sample 3 Iron interferes with the test The interferences will be eliminated as follows Determine the amount of total iron present in the water sample Add an iron standard solution with the same concentration to the vial point 1 instead of deionised water to produce the blank 4 Conversion mg l NH4 mg l N x 1 29 m...

Страница 32: ... stirring rod and dissolve the tablet 6 Add one BORON No 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 20 minutes After th...

Страница 33: ...n the correct sequence 2 The sample solution should have a pH value between 6 and 7 3 Interferences are prevented by the presence of EDTA in the tablets 4 The rate of colour development depends on the temperature The temperature of the sample must be 20 C 1 C 5 B H3 BO3 ...

Страница 34: ...er and empty it leaving a few drops remaining in the vial 5 Add one DPD No 1 tablet straight from the foil and crush the tablet using a clean stirring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key The result is shown i...

Страница 35: ...oided The analysis must take place immediately after taking the sample 3 The DPD colour development is carried out at a pH value of 6 2 to 6 5 The reagent tablet therefore contains a buffer for the pH adjustment Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the reagent is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 4 Exceeding the measuri...

Страница 36: ...d determination with the arrow keys and Confirm with key 6 3 Chloramine Mono with NH4 without NH4 with NH4 without NH4 Notes 1 Full colour development temperature The reaction periods indicated in the manual refer to a sample temperature between 18 and 20 C Due to the fact that the reaction period is strongly influenced by sample temperature you have to adjust both reaction periods according to th...

Страница 37: ... Powder Pack straight from the foil into the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents approx 20 seconds 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts automatically The result is shown in the displ...

Страница 38: ... foil into the chloramine vial 6 Close the vial tightly with the cap and swirl several times to mix the contents approx 20 seconds 7 Add 1 drop of Vario Free Ammonia Reagent Solu tion into the ammonia vial Note 1 8 Close the vial tightly with the cap and invert several times to mix the contents 9 Place the chloramine vial in the sample chamber making sure that the marks are aligned 10 Press TEST k...

Страница 39: ...ee Ammonia The reading of Monochloramine is shown as mg l Cl2 mg l NH2 Cl mg l N The reading of Ammonia is shown as N Conversion mg l NH4 mg l N 1 29 mg l NH3 mg l N 1 22 Notes 1 Hold the bottle vertically and squeeze slowly 2 To determine the ammonia concentration the difference between the chloramine T1 and the sum of chloramine and ammonia T2 is calculated If T2 exceeds the range limit the foll...

Страница 40: ...ing rod and dissolve the tablet 6 Add one CHLORIDE T2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl gently several times until the tablet is dissolved Note 1 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 2 minutes Af...

Страница 41: ...to bigger sized particles which can cause false readings 2 High concentrations of electrolytes and organic compounds have different effects on the precipitation reaction 3 Ions which also form deposits with Silver nitrate in acidic media such as Bromides Iodides and Thiocyanates interfere with the analysis 4 Highly alkaline water should if necessary be neutralised using Nitric acid before analysis...

Страница 42: ...k 0 02 2 mg l Cl2 The following selection is shown in the display for the differentiated determination of free combined and total Chlorine for the determination of free Chlorine for the determination of total Chlorine Select the desired determination with the arrow keys and Confirm with key Chlorine diff free total diff free total 0 0 1 0 1 1 0 1 1 ...

Страница 43: ... pH 7 before the reagent is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 5 Exceeding the measuring range Concentrations above 10 mg l Chlorine using tablets 4 mg l Chlorine using liquid reagents 2 mg l using powder packs can lead to results showing 0 mg l In this case the water sample must be diluted with water free of Chlorine 10 ml of the diluted sample should be mixed with th...

Страница 44: ...ght from the foil and crush the tablet using a clean stirring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key 10 Remove the vial from the sample chamber 11 Add one DPD No 3 tablet straight from the foil to the same water...

Страница 45: ...ST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically The result is shown in the display in mg l free Chlorine mg l combined Chlorine mg l total Chlorine Notes See page 41 T1 accepted prepare T2 press TEST Countdown 2 00 mg l free Cl mg l comb Cl mg l total Cl ...

Страница 46: ...ty it leaving a few drops remaining in the vial 5 Add one DPD No 1 tablet straight from the foil and crush the tablet using a clean stirring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key The result is shown in the disp...

Страница 47: ...ne DPD No 3 tablet straight from the foil and crush the tablets using a clean stirring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the m...

Страница 48: ...me size by holding the bottle vertically and squeeze slowly 6 drops of DPD 1 buffer solution 2 drops of DPD 1 reagent solution 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times to mix the contents 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key 10 Remove the vial from the sample chamber 11 Add 3 drops o...

Страница 49: ...inished the measurement starts automatically The result is shown in the display in mg l free Chlorine mg l combined Chlorine mg l total Chlorine Notes 1 After use replace the bottle caps securely noting the colour coding 2 Store the reagent bottles in a cool dry place ideally between 6 C and 10 C 3 Also see page 41 T1 accepted prepare T2 press TEST Countdown 2 00 mg l free Cl mg l comb Cl mg l tot...

Страница 50: ...ueeze slowly 6 drops of DPD 1 buffer solution 2 drops of DPD 1 reagent solution 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times to mix the contents 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key The result is shown in the display in mg l free Chlorine Notes free and total Chlorine 1 After use replace...

Страница 51: ...le vertically and squeeze slowly 6 drops of DPD 1 buffer solution 2 drops of DPD 1 reagent solution 3 drops of DPD 3 solution 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times to mix the contents 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 2 minutes After the reaction p...

Страница 52: ...ater sample 6 Close the vial tightly with the cap and swirl several times to mix the contents approx 20 seconds 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key 9 Remove the vial from the sample chamber empty the vial rinse vial and cap several times and then fill the vial with 10 ml of the water sample 10 Add the contents of one VARIO Chlorine TOTAL D...

Страница 53: ...ST key Wait for a reaction period of 3 minutes After the reaction period is finished the measurement starts automatically The result is shown in the display in mg l free Chlorine mg l combined Chlorine mg l total Chlorine Notes See page 41 mg l free Cl mg l comb Cl mg l total Cl T1 accepted prepare T2 press TEST Countdown 3 00 ...

Страница 54: ... vial from the sample chamber 5 Add the contents of one VARIO Chlorine FREE DPD F10 Powder Pack straight from the foil to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents approx 20 seconds 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display in mg l free Chlorine Notes See...

Страница 55: ...orine TOTAL DPD F10 Powder Pack straight from the foil to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents approx 20 seconds 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 3 minutes After the reaction period is finished the measurement starts automatically The result is s...

Страница 56: ...ction is shown in the display for the determination of Chlorine dioxide in the presence of Chlorine for the determination of Chlorine dioxide in the absence of Chlorine Select the desired determination with the arrow keys and Confirm with key Chlorine dioxide with Cl without Cl with Cl without Cl 2 0 1 ...

Страница 57: ...immediately after taking the sample 3 The DPD colour development is carried out at a pH value of 6 2 to 6 5 The reagent tablet therefore contains a buffer for the pH adjustment Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the tablet is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 4 Exceeding the measuring range Concentrations above 19 mg ...

Страница 58: ...rom the foil and crush the tablet using a clean stirring rod 6 Fill a second clean vial with 10 ml of the water sample 7 Add one GLYCINE tablet straight from the foil and crush the tablet using a clean stirring rod 8 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 9 Transfer the contents of the second vial into the prepared vial 10 Close the vial tightly w...

Страница 59: ...D No 3 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 21 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 22 Place the vial in the sample chamber making sure that the marks are aligned 23 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts...

Страница 60: ...y 0 38 mg l ClO2 mg l ClO2 Cl x 0 38 ClO2 Cl ClO2 Chlorine dioxide displayed as Chlorine units ClO2 Cl has its origin in swimming poolwater treatment according to DIN 19643 2 The total Chlorine result given includes the contribution of the Chlorine dioxide as Chlorine reading For true total Chlorine value subtract the Chlorine dioxide as Chlorine reading from the quoted total Chlorine reading 3 Al...

Страница 61: ...ps remaining in the vial 5 Add one DPD No 1 tablet straight from the foil and crush the tablet using a clean stirring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key The result is shown in the display as Chlorine dioxide...

Страница 62: ...ORINE HR KI tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Add one ACIDIFYING GP tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are...

Страница 63: ...61 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 Oxidizing agents interfere as they react in the same way as Chlorine ...

Страница 64: ...reactor at a temperature of 150 C 5 CAUTION The vials are hot Remove the tubes from the heating block and allow them to cool to 60 C or less Mix the contents by carefully inverting each tube several times while still warm Then allow the tubes to cool to ambient temperature before measuring Note 2 6 Place the vial the blank Note 3 4 in the sample chamber making sure that the marks are l aligned 7 P...

Страница 65: ...nts 3 Suspended solids in the vial lead to incorrect measurements For this reason it is important to place the vials carefully in the sample chamber The precipitate at the bottom of the sample should be not suspended 4 Clean the outside of the vials with a towel Finger prints or other marks will be removed 5 Samples can be measured when the Chloride content does not exceed 1000 mg l 6 In exception...

Страница 66: ... reactor at a temperature of 150 C 5 CAUTION The vials are hot Remove the tubes from the heating block and allow them to cool to 60 C or less Mix the contents by carefully inverting each tube several times while still warm Then allow the tubes to cool to ambient temperature before measuring Note 2 6 Place the vial the blank Note 3 4 in the sample chamber making sure that the marks are l aligned 7 ...

Страница 67: ...ct measurements For this reason it is important to place the vials carefully in the sample chamber The precipitate at the bottom of the sample should be not suspended 4 Clean the outside of the vials with a towel Finger prints or other marks will be removed 5 Samples can be measured when the Chloride content does not exceed 1000 mg l 6 In exceptional cases compounds contained in the water cannot b...

Страница 68: ...preheated reactor at a temperature of 150 C 5 CAUTION The vials are hot Remove the tubes from the heating block and allow them to cool to 60 C or less Mix the contents by carefully inverting each tube several times while still warm Then allow the tubes to cool to ambient temperature before measuring Note 2 6 Place the vial the blank Note 3 4 in the sample chamber making sure that the marks are l a...

Страница 69: ...son it is important to place the vials carefully in the sample chamber The precipitate at the bottom of the sample should be not suspended 4 Clean the outside of the vials with a towel Finger prints or other marks will be removed 5 Samples can be measured when the Chloride content does not exceed 10000 mg l 6 In exceptional cases compounds contained in the water cannot be oxidized adequately so re...

Страница 70: ... sample measurement 1 Fill a clean vial 24 mm Ø with 10 ml of the filtrated deionised water from Step A close tightly with the cap 2 Place the vial in the sample chamber making sure that the marks are aligned 3 Press ZERO key 4 Remove the vial from the sample chamber and empty it completely 5 Rinse the vial with the filtrated water sample and fill with 10 ml filtrated water sample from Step B 6 Pl...

Страница 71: ...n 3 The estimated detection limit is 15 mg L Pt 4 Colour may be expressed as apparent or true colour The apparent colour is defined as the colour of a solution due to dissolved substances and suspended particles in the sample This manual describes the determination of true colour by filtration of the water sample To determine the apparent colour non filtrated deionised water and sample are measure...

Страница 72: ...e differentiated determination of free combined and total Copper for the determination of free Copper for the determination of total Copper Select the desired determination with the arrow keys and Confirm with key Note 1 If is displayed at the diffentiated test result see page 262 Copper diff free total diff free total 5 0 1 ...

Страница 73: ... the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key 9 Remove the vial from the sample chamber 10 Add one COPPER No 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 11 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 12 Place the vial i...

Страница 74: ...ve the vial from the sample chamber 5 Add one COPPER No 1 tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display in mg l free Copper pr...

Страница 75: ...m the sample chamber 5 Add one COPPER No 1 tablet and one COPPER No 2 tablet straight from the foil to the water sample and crush the tablets using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display in mg l...

Страница 76: ...e contents of one VARIO Cu 1 F10 Powder Pack straight from the foil to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents Note 3 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically The result is...

Страница 77: ...undissolved powder 4 Interferences Cyanide CN Cyanide prevents full colour development Add 0 2 ml Formaldehyde to 10 ml water sample and wait for a reaction time of 4 minutes Cyanide is masked After this perform test as described Multiply the result by 1 02 to correct the sample dilution by Formaldehyde Silver Ag If a turbidity remains and turns black silver interference is likely Add 10 drops of ...

Страница 78: ...ontents 6 Add two level spoons No 4 grey of Cyanide 12 replace the cap tightly and invert the vial several times to mix the contents 7 Fill the vial with drops of the same size by holding the bottle vertically and squeeze slowly 3 drops of Cyanide 13 8 Close the vial tightly with the cap and invert several times to mix the contents 9 Place the vial in the sample chamber making sure that the marks ...

Страница 79: ...e destroyed by Chlorine are determined by this test 2 In the present of Thiocyanate heavy metal complexes colorants or aromatic amines the cyanide must be separated out by distillation before analysis is performed 3 Store the reagents in closed containers at a temperature of 15 C to 25 C ...

Страница 80: ... Remove the vial from the sample chamber 5 Add one CYANURIC ACID tablet straight from the foil to the prepared water sample and crush the tab let using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved Note 2 3 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the disp...

Страница 81: ... Cyanuric acid 2 If Cyanuric acid is present a cloudy solution will occur Small single particles are not necessarily caused by Cyanuric acid 3 Dissolve the tablet completely therefore swirl the vial approx 1 minute Un dissolved particles of the tablet can cause results that are too high ...

Страница 82: ...6 drops 0 25ml of DEHA solution 6 Close the vial tightly with the cap and swirl several times to mix the contents 7 Add one DEHA tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 8 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 9 Place the vial in the sample chamber making sure that the marks are aligne...

Страница 83: ...res at all concentrations Repeat the test procedure but without adding the DEHA solution If the displayed result is above 20 μg l subtract this value from the DEHA test result Substances which reduce Iron III interfere Substances which complex iron strongly may interfere also Substances which may interfere when present in concentrations at Borate as Na2 B4 O7 500 mg l Cobalt 0 025 mg l Copper 8 0 ...

Страница 84: ...wirl several times to mix the contents 5 Add 0 20 ml VARIO DEHA 2 Rgt Solution to each vial Note 4 6 Close the vials tightly with the caps and swirl several times to mix the contents 7 Press key Wait for a reaction period of 10 minutes Note 5 After the reaction period is finished proceed as follows 8 Place the vial the blank in the sample chamber making sure that the marks are aligned 9 Press ZERO...

Страница 85: ...s 6 Interferences Iron II interferes at all concentrations Repeat the test procedure but without adding the VARIO DEHA Rgt 2 solution If the displayed result is above 20 μg l subtract this value from the DEHA test result Substances which reduce Iron III interfere Substances which complex iron strongly may interfere also Substances who may interfere when present in concentrations at Borate as Na2 B...

Страница 86: ...ned 3 Press ZERO key 4 Remove the vial from the sample chamber 5 Add exactly 2 ml SPADNS reagent solution Note 4 to the water sample Caution Vial is filled up to the top Note 8 6 Close the vial tightly with the cap and swirl several times to mix the contents 7 Place the vial in the sample chamber making sure that the marks are aligned Press TEST key The result is shown in the display in mg l Fluor...

Страница 87: ...samples to be tested should have the same temperature 1 C 4 As the test result is highly dependent on exact sample and reagent volumes the sample and reagent volumes should always be metered by using a 10 ml resp 2 ml volumetric pipette class A 5 The accuracy of the test methods decreases above a level of 1 2 mg l Fluoride Although the results are sufficiently accurate for most applications even m...

Страница 88: ...he marks are aligned 5 Press ZERO key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically 6 Remove the vial from the sample chamber 7 Add 2 ml of the water sample to the prepared vial Caution Vial is filled up to the top Note 4 8 Close the vial tightly with the cap and swirl several times 5x to mix the contents 9 Place the vial in the ...

Страница 89: ...lerance of the method is increasing with higher concentrations When diluting samples this should be take into account always measuring in the first third of the range 3 This method was developed from a volumetric procedure for the determination of calcium Due to undefined conditions the deviations from the standardised method may be greater 4 It is convenient to use special vials with larger volum...

Страница 90: ...g rod and dissolve the tablet completely 6 Add one CALCIO H No 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl gently several times until the tablet is completely dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period ...

Страница 91: ...e must be taken for the test 4 This method was developed from a volumetric procedure for the determination of Calcium Hardness Due to undefined conditions the deviations from the standardised method may be greater 5 The tolerance of the method is increasing with higher concentrations When diluting samples this should be taken in account always measuring in the first third of the range 6 Interferen...

Страница 92: ...et straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts automatically The ...

Страница 93: ...amples must be adjusted between pH 4 and pH 10 before the tablet is added use 1 mol l Hydrochloric acid resp 1mol l Sodium hydroxide 2 Conversion table mg l CaCO3 dH fH eH 1 mg l CaCO3 0 056 0 10 0 07 1 dH 17 8 1 78 1 25 1 fH 10 0 0 56 0 70 1 eH 14 3 0 80 1 43 3 CaCO3 dH eH fH aH ...

Страница 94: ...HARDCHECK P tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts au...

Страница 95: ...amples must be adjusted between pH 4 and pH 10 before the tablet is added use 1 mol l Hydrochloric acid resp 1mol l Sodium hydroxide 2 Conversion table mg l CaCO3 dH fH eH 1 mg l CaCO3 0 056 0 10 0 07 1 dH 17 8 1 78 1 25 1 fH 10 0 0 56 0 70 1 eH 14 3 0 80 1 43 3 CaCO3 dH eH fH aH ...

Страница 96: ...st powder Note 3 to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents 7 Press key Wait for a reaction period of 10 minutes After the reaction period is finished proceed as follows 8 The slight turbidity that occurs when the reagent is added must be removed by filtration Note 4 9 Place the vial in the sample chamber making sure that the marks are ali...

Страница 97: ...level spoon 4 Qualitative folded filter papers for medium precipitates are recommended 5 In order to check whether the reagent has aged if it has been stored for a lengthy period perform the test as described above using tap water If the result is above the detection limit of 0 05 mg l you should only use the reagent with reservations as there may be a major deviation in results 6 There is an opti...

Страница 98: ...he marks are aligned 5 Press ZERO key 6 Remove the vial from the sample chamber 7 Fill the second clean vial with 10 ml of the water sample this is the sample 8 Add 1 ml VARIO Hydra 2 Rgt Solution into the vial 9 Close the vial tightly with the cap and swirl several times to mix the contents 10 Place the vial the blank in the sample chamber making sure that the marks are aligned 11 Press TEST key ...

Страница 99: ...ble that the test result increases by 20 Morpholine does not interfere up to 10 mg l Highly coloured or turbid samples Mix 1 part deionised water with 1 part household bleach Add 1 drop of this mixture into 25 ml water sample and mix Use 10 ml prepared sample in place of deionised water in point 1 Note at point 7 use the unprepared water sample Principle Hydrazine is oxidised by household bleach C...

Страница 100: ...g the vial against the side of the sample container The Vacu vial breaks at the neck and the vial fills automatically A small volume of inert gas remains in the Vacu vial 6 Mix the contents of the Vacu vial by inverting it several times allowing the bubble to move from one end to the other Dry the outside of the vial 7 Place the Vacu vial in the sample chamber 8 Press TEST key Wait for a reaction ...

Страница 101: ...d the original test instruction and the MSDS delivered with the test before performing the test MSDS also available at www chemetrics com 3 Vacu vials is a registered trade mark of the company CHEMetrics Inc Calverton U S A 4 There is an option to change the unit from mg l to μg l mg l µg l ...

Страница 102: ... HYDROGENPEROXIDE LR tablet straight from the foil and crush the tablet using a clean stir ring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finishe...

Страница 103: ...ed The analysis must take place immediately after taking the sample 3 The DPD colour development is carried out at a pH value of 6 2 to 6 5 The reagent tablet therefore contains a buffer for the pH adjustment Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the tablet is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 4 Exceeding the measuring r...

Страница 104: ...empty the vial leaving a view drops in 5 Add one DPD No 1 tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key The result is shown ...

Страница 105: ...103 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 Oxidizing reagents such as Chlorine Bromine etc interfere as they react in the same way as Iodine ...

Страница 106: ...f all dissolved iron and most undissolved forms of iron Iron total with Vario Powder Pack 0 02 1 8 mg l Fe Determination of all dissolved iron and most undissolved forms of iron most undissolved iron oxides are recovered by the reagent This information refers to analysis of the water sample without digestion Further information can be found in the method notes 2 0 2 2 2 2 2 3 2 ...

Страница 107: ...l of this pre treated solution is used for the following analysis Perform as described by the selected test method 2 Water which has been treated with organic compounds like corrosion inhibitors must be oxidised where necessary to break down the iron Therefore add 1 ml concentrated sulfuric acid and 1 ml concentrated nitric acid to 100 ml water sample and boil to approx half volume After cooling d...

Страница 108: ...t straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts automatically The r...

Страница 109: ...nes the total dissolved Iron as Fe2 and Fe3 2 The IRON II LR tablet is used for differentiation as described above instead of the IRON LR tablet 3 For the determination of total dissolved and undissolved iron digestion is required An example is described on page 105 ...

Страница 110: ...ents of one Vario Ferro F10 Powder Pack straight from the foil to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents Note 4 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 3 minutes Note 5 After the reaction period is finished the measurement starts automatically The result ...

Страница 111: ...requires prior digestion use mild vigorous or Digesdahl digestion e g for digestion with acid see page 105 3 Very strong alkaline or acidic water samples must be adjusted to a pH value between 3 and 5 before analysis 4 Accuracy is not affected by undissolved powder 5 Water samples containing visible rust should be allowed to react for at least five minutes ...

Страница 112: ... vial 4 Close the vials tightly with the caps and swirl several times to mix the contents 5 Press key Wait for a reaction period of 3 minutes After the reaction period is finished proceed as follows 6 Place the vial the blank in the sample chamber making sure that the marks are aligned 7 Press ZERO key 8 Remove the vial from the sample chamber 9 Place the vial the sample in the sample chamber ma k...

Страница 113: ...between pH 3 and pH 8 before the reagent is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 4 Interferences When interferences occur colour development is inhibited or a precipitate is formed The values below refer to a standard with an iron concentration of 0 5 mg l The following substances do not interfere when present up to the levels given Substance no inerference to Cadmium 4 ...

Страница 114: ...n stirring rod and dissolve the tablet 6 Add one MANGANESE LR 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 5 minutes Afte...

Страница 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...

Страница 116: ...t solution 6 Close the vials tightly with the caps and swirl several times to mix the contents 7 Fill each vial with drops of the same size by holding the bottle vertically and squeeze slowly 21 drops of PAN Indicator solution 8 Close the vials tightly with the caps and swirl several times to mix the contents 9 Press key Wait for a reaction period of 2 minutes Note 4 After the reaction period is f...

Страница 117: ...dding the Vario Ascorbic Acid powder pack add additionally 10 drops of Rochelle Salt Solution 3 After addition of the reagent solution Alkaline Cyanide a cloudy or turbid solution may form in some water samples The turbidity should disappear after point 7 4 Water samples containing more than 5 mg l iron should be allowed to react for at least 10 minutes 5 Conversion mg l MnO4 mg l Mn x 2 17 6 Mn M...

Страница 118: ... Close the vial tightly with the cap and swirl several times to mix the contents 7 Add the contents of one VARIO Sodium Periodate F10 Powder Pack straight from the foil to the same water sample 8 Close the vial tightly with the cap and swirl several times to mix the contents 9 Place the vial in the sample chamber making sure that the marks are aligned 10 Press TEST key Wait for a reaction period o...

Страница 119: ...apacity of the reagents and requires sample pre treatment If samples were acidified for storing adjust the pH between 4 and 5 with 5 mol l 5 N Sodium hydroxide before test Do not exceed pH 5 as manganese may precipitate 3 Interferences Interfering substance Interference level Calcium greater than 700 mg l Chloride greater than 70 000 mg l Iron greater than 5 mg l Magnesium greater than 100 000 mg ...

Страница 120: ...traight from the foil to the water sample and crush the tablet using a clean stirring rod 7 Add one MOLYBDATE HR No 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 8 Dissolve the tablets using a clean stirring rod 9 Rinse out the vial with the prepared water sample and then fill to the 10 ml mark 10 Close the vial tightly with the cap 11 Pla...

Страница 121: ...must be added in the correct sequence 2 Under test conditions pH 3 8 3 9 iron does not interfere nor do other metals at levels likely to be found in industrial water systems 3 Conversions mg l Mo mg l MoO4 x 0 6 mg l Na2 MoO6 mg l MoO4 x 1 3 4 MoO4 Mo Na2 MoO4 ...

Страница 122: ...ntents of one Vario Molybdenum HR 2 F10 Powder Pack straight from the foil to the same water sample 8 Close the vial tightly with the cap and swirl several times to mix the contents 9 Add the contents of one Vario Molybdenum HR 3 F10 Powder Pack straight from the foil to the same water sample 10 Close the vial tightly with the cap and swirl several times to mix the contents 11 Place the vial in th...

Страница 123: ...early 7 with 1 mol l Nitric acid or 1 mol l Sodium hydroxide 3 Concentrations above 10 mg l Cu causes too high test values if the reaction time of 5 minutes is increased So it is very important to perform the test procedure as described 4 Substances which may interfere when present in concentrations at Aluminium 50 mg l Chromium 1000 mg l Iron 50 mg l Nickel 50 mg l Nitrite all levels 5 MoO4 Mo Na...

Страница 124: ... tightly with the caps and invert gently several times 10 x to mix the contents Note 1 5 Press key Wait for a reaction period of 5 minutes 6 After the reaction period is finished proceed as follows 7 Place the vial the blank in the sample chamber making sure that the marks are l aligned 8 Press ZERO key 9 Remove the vial from the sample chamber 10 Place the vial the sample in the sample chamber ma...

Страница 125: ...123 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 Some solids may not dissolve 2 Conversion mg l NO3 mg l N x 4 43 3 N NO3 ...

Страница 126: ...traight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 10 minutes After the reaction period is finished the measurement starts automatically The res...

Страница 127: ...y III Iron III Lead Mercury I Silver Chloroplatinate Metavanadate and Bismuth Copper II ions may cause lower test results as they accelerate the decomposition of the Diazonium salt It is unlikely in practice that these interfering ions will occur in such high concentrations that they cause significant reading errors 2 Conversion mg l NO2 mg l N x 3 29 3 N NO2 ...

Страница 128: ... contents of one VARIO Nitri 3 Powder Pack straight from the foil to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 20 minutes After the reaction period is finished the measurement starts automatically The result is shown i...

Страница 129: ...ric Bismuth Chloroplatinate Ferric Lead Mercurous Metavanadate Silver ions interfere by causing precipitation In samples with very high concentrations of Nitrate 100 mg L N a small amount of Nitrite will be found Such high levels of Nitrate appear to undergo a slight amount of reduction to Nitrite either spontaneously or during the reaction time of the test 2 N NO2 ...

Страница 130: ...ne Vario TN Reagent A Powder Pack to each vial Note 2 8 Close the vials with the caps and shake to mix the contents at least 15 seconds 9 Press key Wait for a reaction period of 3 minutes After the reaction period is finished proceed as follows 10 Open the digestion vials and add the contents of one Vario TN Reagent B Powder Pack to each vial Note 2 11 Close the vials with the caps and shake to mi...

Страница 131: ...xactly 30 minutes 8 The reagent will not completely dissolve 9 Hold the vial in a vertical position with the cap pointing up Turn the vial upside down Wait for all of the solution to flow down to the cap Return the vial to the upright position Wait for all the solution to flow to the bottom of the vial This process is one inversion 10 inversions approx 30 seconds 10 After zero calibration with the...

Страница 132: ...ario TN Reagent A Powder Pack to each vial Note 2 8 Close the vials with the caps and shake to mix the contents at least 15 seconds 9 Press key Wait for a reaction period of 3 minutes After the reaction period is finished proceed as follows 10 Open the digestion vials and add the contents of one Vario TN Reagent B Powder Pack to each vial Note 2 11 Close the vials with the caps and shake to mix th...

Страница 133: ...ly 30 minutes 8 The reagent will not completely dissolve 9 Hold the vial in a vertical position with the cap pointing up Turn the vial upside down Wait for all of the solution to flow down to the cap Return the vial to the upright position Wait for all the solution to flow to the bottom of the vial This process is one inversion 10 inversions approx 30 seconds 10 After zero calibration with the bla...

Страница 134: ...traight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically The resu...

Страница 135: ...ve Oxygen is a synonym for a common disinfectant based on Oxygen in Swimming Pool Treatment 1 When preparing the sample the lost of Oxygen e g by pipetting or shaking must be avoided 2 The analysis must take place immediately after taking the sample ...

Страница 136: ... lower edge of the sample contai ner The Vacu vial breaks at the neck and the vial fills automatically A small volume of inert gas remains in the Vacu vial 6 Remove the Vacu vial point downwards from the sample container immediately As the contents of the vial has a higher density than water it is important to remove the vial from the sam ple container within 5 seconds to prevent any loss of reage...

Страница 137: ...d is adapted from CHEMetrics 2 Read the original test instruction and the MSDS delivered with the test before performing the test MSDS also available at www chemetrics com 3 Vacu vials is a registered trade mark of the company CHEMetrics Inc Calverton U S A ...

Страница 138: ...following selection is shown in the display for the determination of Ozone in the presence of Chlorine for the determination of Ozone in the absence of Chlorine Select the desired method with the arrow keys and Confirm with key Ozon with Cl without Cl with Cl without Cl 0 0 3 ...

Страница 139: ...er for the pH adjustment Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the tablet is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 4 Turbidity can lead to errors The use of the DPD No 1 tablet in samples with high Calcium ion content and or high conductivity can lead to turbidity of the sample and therefore incorrect measurements it is not ...

Страница 140: ...ng rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically 10 Remove the vial from the sample chamber empty the...

Страница 141: ...ts of the second vial into the prepared vial 16 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 17 Place the vial in the sample chamber making sure that the marks are aligned 18 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically The result is shown in the display in mg l Ozone...

Страница 142: ...ablet and one DPD No 3 tablet straight from the foil and crush the tablets using a clean stirring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is fin...

Страница 143: ...141 MaxiDirect_2c 03 2010 ...

Страница 144: ...Remove the vial from the sample chamber 5 Add one PHMB PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display in mg l PHMB p...

Страница 145: ...tirring rods with a laboratory detergent see chapter 1 2 2 Cleaning of vials and accessories for analysis Rinse vials and caps thoroughly with tap water and then with deionised water 3 The test result is influenced by Hardness and Total Alkalinity The calibration of this method was done using water with the following concentration Ca Hardness 200 mg l CaCO3 Total Alkalinity 120 mg l CaCO3 ...

Страница 146: ... PO4 Determination of ortho Phosphate ions Phosphat 1 ortho with Vacu vials 5 40 mg l PO4 Determination of ortho Phosphate ions Phosphat 2 ortho with Vacu vials 0 05 5 mg l PO4 Determination of ortho Phosphate ions Phosphate acid hydrolizable with Vario Tube Test 0 02 1 6 mg l P Determination of ortho Phosphate ions condensed inorganic Phosphates Phosphate total with Vario Tube Test 0 02 1 1 mg l ...

Страница 147: ...21 and 327 the ortho Phosphate ions react with the Vanadate molybdate reagent under acid conditions to form a yellow coloured product Notes only for tube tests and tests with powder packs 323 324 325 326 1 Application for water wastewater and seawater 2 Highly buffered samples or samples with extreme pH values should be adjusted between pH 2 and pH 10 before analysis with 1 mol l Hydrochloric acid...

Страница 148: ...using a clean stirring rod 6 Add one PHOSPHATE No 2 LR tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 10 minutes After the reac...

Страница 149: ...sequence 3 The test sample should have a pH Value between 6 and 7 4 Interferences Higher concentrations of Cu Ni Cr III V V and W VI interfere due to their colour Silicates do not interfere masked by Citric acid in the tablets 5 see also page 145 6 Conversion mg l P mg l PO4 x 0 33 mg l P2 O5 mg l PO4 x 0 75 7 PO4 P P2 O5 ...

Страница 150: ... using a clean stirring rod 6 Add one PHOSPHATE HR P2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 10 minutes After the rea...

Страница 151: ... samples under 5 mg l PO4 it is reccommended to analyse the water sample with method 320 Posphate LR ortho with Tablet 2 Only ortho Phosphate ions react 3 see also page 145 4 Conversions mg l P mg l PO4 x 0 33 mg l P2 O5 mg l PO4 x 0 75 5 PO4 P P2 O5 ...

Страница 152: ...ARIO Phosphate Rgt F10 Powder Pack straight from the foil to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents approx 10 15 sec Note 1 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically The re...

Страница 153: ...151 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 The reagent does not dissolve completely 2 see also page 145 3 Conversions mg l P mg l PO4 x 0 33 mg l P2 O5 mg l PO4 x 0 75 4 PO4 P P2 O5 ...

Страница 154: ...of one VARIO Phosphate Rgt F10 Powder Pack straight from the foil to the water sample Note 1 6 Close the vial tightly with the cap and swirl several times to mix the contents approx 10 15 sec Note 2 7 Place the vial in the sample chamber making sure that the l marks are aligned 8 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts aut...

Страница 155: ...Direct_2c 03 2010 1 1 Methods Notes 1 Use a funnel to add the reagent 2 The reagent does not dissolve completely 3 see also page 145 4 Conversions mg l P mg l PO4 x 0 33 mg l P2 O5 mg l PO4 x 0 75 5 PO4 P P2 O5 ...

Страница 156: ...gainst the side of the sample container The Vacu vial breaks at the neck and the vial fills au tomatically A small volume of inert gas remains in the Vacu vial 6 Mix the contents of the Vacu vial by inverting it several times allowing the bubble to move from one end to the other Dry the outside of the vial 7 Place the Vacu vial in the sample chamber 8 Press TEST key Wait for a reaction period of 5...

Страница 157: ...instruction and the MSDS delivered with the test before performing the test MSDS also available at www chemetrics com 3 Vacu vials is a registered trade mark of the company CHEMetrics Inc Calverton U S A 4 Only ortho Phosphate ions react 5 Sulfide Thiosulfate and Thiocyanate cause low test results 6 PO4 P P2 O5 ...

Страница 158: ...irl several times to mix the contents 7 Place one Vacu vial in the sample container Snap the tip by pressing the vial against the side of the sample container The Vacu vial breaks at the neck and the vial fills automatically A small volume of inert gas remains in the Vacu vial 8 Mix the contents of the Vacu vial by inverting it several times allowing the bubble to move from one end to the other Dr...

Страница 159: ...instruction and the MSDS delivered with the test before performing the test MSDS also available at www chemetrics com 3 Vacu vials is a registered trade mark of the company CHEMetrics Inc Calverton U S A 4 Only ortho Phosphate ions react 5 Sulfide Thiosulfate and Thiocyanate cause low test results 6 PO4 P P2 O5 ...

Страница 160: ...o the vial 6 Close the vial with the cap and invert gently several times to mix the contents 7 Place the vial in the sample chamber making sure that the l marks are aligned 8 Press ZERO key 9 Remove the vial from the sample chamber 10 Add the contents of one VARIO Phosphate Rgt F10 Powder Pack straight from the foil to the vial Note 2 11 Close the vial tightly with the cap and swirl several times ...

Страница 161: ... safety precautions and a good lab technique should be used during the whole procedure 2 Use a funnel to add the reagent 3 The reagent does not dissolve completely 4 see also page 145 5 Conversions mg l PO4 mg l P x 3 07 mg l P2 O5 mg l P x 2 29 6 PO4 P P2 O5 ...

Страница 162: ... vial and add 2 ml 1 54 N Sodium Hydroxide Solution to the vial 7 Close the vial with the cap and invert gently several times to mix the contents 8 Place the vial in the sample chamber making sure that the l marks are aligned 9 Press ZERO key 10 Remove the vial from the sample chamber 11 Add the contents of one VARIO Phosphate Rgt F10 Powder Pack straight from the foil to the vial Note 2 12 Close ...

Страница 163: ... safety precautions and a good lab technique should be used during the whole procedure 2 Use a funnel to add the reagent 3 The reagent does not dissolve completely 4 see also page 145 5 Conversions mg l PO4 mg l P x 3 07 mg l P2 O5 mg l P x 2 29 6 P PO4 P2 O5 ...

Страница 164: ...ersulfate F10 Powder Pack straight from the foil to the digestion vial 6 Close the digestion vial tightly with the cap and swirl until the reagent is dissolved completely 7 Insert the UV lamp into the digestion vial Note 3 4 5 CAUTION Wear UV safety goggles 8 Switch the UV lamp on and wait for a reaction period of 10 minutes 9 After the reaction period is finished switch the UV lamp off and remove...

Страница 165: ...actual phosphonate concentration using the appropriate factor from table 2 Notes 1 Rinse all glassware with 1 1 Hydrochloric acid first and then rinse with deionised water Do not use detergents with phosphates 2 During UV digestion Phosphonates are converted to Ortho Phosphates This step is normally completed in 10 minutes High organic loaded samples or a weak lamp can cause incomplete phosphate c...

Страница 166: ...ge mg L Phosphonate Sample volume in ml Factor 0 2 5 50 0 1 0 5 0 25 0 2 0 12 5 10 0 5 0 25 5 1 0 0 125 1 5 0 Table 2 Phosphonate type Conversion factor for active phosphonate PBTC 2 840 NTP 1 050 HEDPA 1 085 EDTMPA 1 148 HMDTMPA 1 295 DETPMPA 1 207 HPA 1 490 ...

Страница 167: ...ide 100 mg l Calcium 5000 mg l CDTA 100 mg l Chloride 5000 mg l Chromate 100 mg l Copper 100 mg l Cyanide 100 mg l increase the UV digestion to 30 minutes Diethanoldithiocarbamate 50 mg l EDTA 100 mg l Iron 200 mg l Nitrate 200 mg l NTA 250 mg l Orthophosphate 15 mg l Phosphite and organophosphorus compounds reacts quantitatively Meta and Polyphosphates do not interfere Silica 500 mg l Silicate 10...

Страница 168: ...ial from the sample chamber 5 Add one BROMOCRESOLPURPLE PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display as pH value p...

Страница 169: ...lausibility test pH meter is recommended 3 The accuracy of the colorimetric determination of pH values depends on various boundary conditions buffer capacity of the sample salt contents etc 4 Salt error Correction of test results average values for samples with salt contents of Indicator Salt content Bromcresolpurple 1 molar 0 26 2 molar 0 33 3 molar 0 31 The values of Parson and Douglas 1926 are ...

Страница 170: ... vial from the sample chamber 5 Add one PHENOL RED PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display as pH value prepar...

Страница 171: ... produce results inside the measuring range A plausibility test pH meter is recommended 4 The accuracy of the colorimetric determination of pH values depends on various boundary conditions buffer capacity of the sample salt contents etc 5 Salt error Correction of test results average values for samples with salt contents of Indicator Salt content Phenol red 1 molar 0 21 2 molar 0 26 3 molar 0 29 T...

Страница 172: ... Remove the vial from the sample chamber 5 Fill the vial with drops of the same size by holding the bottle vertically and squeeze slowly 6 drops of PHENOL RED solution 6 Close the vial tightly with the cap and swirl several times to mix the contents 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display as pH value prepare ...

Страница 173: ...stal of Sodiumthiosulfate Na2 S2 O3 5 H2 O to the sample before adding the PHENOL RED solution PHENOL RED tablets already contain Thiosulfate 2 Due to differing drop sizes results can show a discrepancy in accuracy by comparison with tablets This can be minimised by using a pipette 0 18 ml PHENOLRED solution is equivalent to 6 drops 3 After use replace the bottle cap securely 4 Store the reagent i...

Страница 174: ...he vial from the sample chamber 5 Add one THYMOLBLUE PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display as pH value prep...

Страница 175: ...ausibility test pH meter is recommended 3 The accuracy of the colorimetric determination of pH values depends on various boundary conditions buffer capacity of the sample salt contents etc 4 Salt error Correction of test results average values for samples with salt contents of Indicator Salt content Thymolblue 1 molar 0 22 2 molar 0 29 3 molar 0 34 The values of Parson and Douglas 1926 are based o...

Страница 176: ...ve the vial from the sample chamber 5 Add one Potassium T tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display in mg l Potassium prep...

Страница 177: ...175 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 If Potassium is present a cloudy solution will appear Single particles are not necessarily caused by Potassium ...

Страница 178: ... minutes After the reaction period is finished proceed as follows 8 Add one SILICA PR tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 9 Add one SILICA No 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 10 Close the cap tightly and swirl several times until the tablets are dissolved 11 Pl...

Страница 179: ...tablets must be added in the correct sequence 2 Phosphate ions do not interfere under the given reaction conditions 3 If Phosphate is known to be absent the addition of the SILICA PR tablet may be omitted 4 Conversion mg l Si mg l SiO2 x 0 47 5 SiO2 Si ...

Страница 180: ...s finished proceed as follows 5 Add the contents of one Vario Silica Citric Acid F10 Powder Pack straight from the foil into each vial 6 Close the vials tightly with the caps and swirl several times to mix the contents 7 Press key Wait for a reaction period of 1 minute Note 3 After the reaction period is finished proceed as follows 8 Place the vial the blank in the sample chamber making sure that ...

Страница 181: ...minutes are required 3 The given reaction time of 1 minute refers to a water sample temperature of 20 C At 30 C a reaction time of 30 seconds at 10 C a reaction time of 2 minutes are required 4 Interferences Substance Interference Iron large amounts interfere Phosphate does not interfere at concentrations less than 50 mg l PO4 at 60 mg l PO4 the interference is approx 2 at 75 mg l PO4 the interfer...

Страница 182: ... the foil to the same water sample Note 2 8 Close the vial tightly with the cap and swirl several times to mix the contents 9 Press key Wait for a reaction period of 10 minutes After the reaction period is finished proceed as follows 10 Add the contents of one Vario Silica Citric Acid F10 Powder Pack straight from the foil to the water sample Note 3 11 Close the vial tightly with the cap and swirl...

Страница 183: ...less than 50 mg l PO4 at 60 mg l PO4 the interference is approx 2 at 75 mg l PO4 the interference is approx 11 Sulfide interferes at all levels Occasionally water samples contain forms of silica which reacts very slowly with Molybdate The nature of these forms is not known A pre treatment with Sodium hydrogencarbonate and then with Sulfuric Acid will make these forms reactive to Molybdate pre trea...

Страница 184: ...a 100 ml beaker and dilute to the 100 ml mark with chlorine free water Mix thoroughly Performing test procedure 1 Fill a clean vial 24 mm Ø with 10 ml of the prepared water sample close tightly with the cap 2 Place the vial in the sample chamber making sure that the marks are aligned 3 Press ZERO key 4 Remove the vial from the sample chamber 5 Add one CHLORINE HR KI tablet straight from the foil t...

Страница 185: ...ite The material has a very strong alkalinity and can cause corrosion Contact with eyes skin and clothes etc has to be avoided Refer to the detailed information the producer supplied with the product 2 The tablets must be added in the correct sequence 3 This method provides a fast and simple test The test can be performed on site but the result will not be as precise as a laboratory method 4 By st...

Страница 186: ...ove the vial from the sample chamber 5 Add one SULFATE T tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display in mg l Sulfate prepare...

Страница 187: ...185 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 If Sulfate is present a cloudy solution will appear ...

Страница 188: ...ntents of one VARIO Sulpha 4 F10 Powder Pack straight from the foil to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts automatically The result is shown ...

Страница 189: ...187 MaxiDirect_2c 03 2010 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will appear ...

Страница 190: ...n stirring rod and dissolve the tablet 6 Add one SULFIDE No 2 tablet to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 10 minutes After the reaction period is...

Страница 191: ...ich react with DPD do not interfere with the test 3 To avoid loss of Sulfide collect the sample carefully with a minimum of aeration It is essential to test the sample immediately after collection 4 The sample temperature should be 20 C A different temperature can lead to higher or lower results 5 Conversion H2 S mg l S x 1 06 6 S H2 S ...

Страница 192: ...straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts automatically The res...

Страница 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...

Страница 194: ...le chamber making sure that the marks are aligned 3 Press ZERO key 4 Remove the vial from the sample chamber and empty the vial completely 5 Stir the blended water sample Immediately rinse the vial with the water sample and fill with 10 ml water sample 6 Place the vial in the sample chamber making sure that the marks are aligned 7 Press TEST key The result is shown in the display in mg l TSS Total...

Страница 195: ...imetric determination of a water sample The result can be used to calibrate the photometer with the same water sample 3 The estimated detection limit is 20 mg L TSS 4 Collect water samples in clean plastic or glass bottles and analyse the water sample as soon as possible It is possible to store the sample at 4 C for 7 days Before measurement warm up the sample to the temperature at collection time...

Страница 196: ...ove the vial from the sample chamber and empty the vial completely 5 Stir the water sample Immediately rinse the vial with the water sample and fill with 10 ml water sample 6 Close the vial tightly with the cap and swirl gently several times 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display in FAU prepare Zero press ZE...

Страница 197: ...er samples in clean plastic or glass bottles and analyse the water sample as soon as possible It is possible to store the sample at 4 C for 48 hours Before measurement warm up the sample to the temperature at collection time Temperature differences between measurement and sample collection can effect the turbidity of the sample 4 Colour interferes if light is absorbed at 530 nm For strong coloured...

Страница 198: ... the cap and swirl several times to mix the contents 7 Add 1 drop of Urea Reagent 2 Urease to the same water sample Note 9 8 Close the vial tightly with the cap and swirl several times to mix the contents 9 Press key Wait for a reaction period of 5 minutes After the reaction period is finished proceed as follows 10 Add one AMMONIA No 1 tablet straight from the foil to the prepared water sample and...

Страница 199: ...e measuring range In this case the water sample should be diluted with Urea free water and remeasured 4 The tablets must be added in the correct sequence 5 The AMMONIA No 1 tablet will only dissolve completely after the AMMONIA No 2 tablet has been added 6 Do not store reagent 1 Urease below 10 C granulation is possible Store reagent 2 Urease in the refrigerator at a temperature of 4 C to 8 C 7 Am...

Страница 200: ...aligned 5 Press ZERO key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts automatically 6 Remove the vial from the sample chamber 7 Add one EDTA tablet straight from the foil to the prepared vial and crush the tablet using a clean stirring rod 8 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 9 Place the ...

Страница 201: ...the case of high levels of residual chlorine perform the analysis with a dechlorinated water sample To dechlorinate add one DECHLOR tablet to the water sample point 1 Crush and mix to dissolve the tablet Then add the COPPER ZINC LR tablet point 2 and continue with the test procedure as described above ...

Страница 202: ...il without touching them with the fingers Liquid reagents Add drops of the same size to the water sample by holding the bottle vertically and squeezing slowly After use replace the bottle caps securely noting the colour coding Note recommendation for storage e g cool and dry Powder Packs Vacu vials from CHEMetrics Vacu vials should be stored in the dark and at room temperature For further informat...

Страница 203: ...ughly with deionised water 1 2 3 Guidelines for photometric measurements 1 Vials caps and stirring rods should be cleaned thoroughly after each analysis to prevent interferences Even minor reagent residues can cause errors in the test result 2 The outside of the vial must be clean and dry before starting the analysis Clean the outside of the vials with a towel Fingerprints or other marks will be r...

Страница 204: ... housing it can destroy electronic components and cause corrosion 8 Contamination of the lens in the sample chamber can result in errors Check at regular intervals and if necessary clean the light entry surfaces of the sample chamber using a moist cloth or cotton buds 9 Large temperature differences between the instrument and the environment can lead to errors e g due to the formation of condensat...

Страница 205: ...thods Caution 1 Dilution decreases accuracy 2 Do not dilute water samples for measurement of pH values This will lead to incorrect test results If Overrange is displayed use another instrument e g pH meter 1 2 5 Correcting for volume additions If a larger volume of acid or base is used to pre adjust the pH value a volume correction of the displayed result is necessary Example For adjusting the pH ...

Страница 206: ...204 MaxiDirect_2c 03 2010 ...

Страница 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...

Страница 208: ...s If the change time exceeds 2 minutes all stored data and settings are lost Recommendation for replacement a screwdriver and new batteries must be available 2 1 3 Replacement of batteries See chapter 2 1 2 Saving data important notes before replacing batteries 1 Switch the instrument off 2 If necessary remove vial from the sample chamber 3 Place the instrument upside down on a clean and even surf...

Страница 209: ...ompartment cover B must be fixed with the four screws 2 1 4 Instrument explosion drawing A screws B battery compartment cover C notch D batteries 4 batteries AA LR6 E seal ring F instrument back D batteries E seal ring C notch F instrument back B battery compartment cover A screws A A L R 6 A A L R 6 A A L R 6 A A L R 6 ...

Страница 210: ...208 MaxiDirect_2c 03 2010 ...

Страница 211: ...s or previous menu Function key description in the text if key available Function key description in the text if key available Function key description in the text if key available Confirming Menu of photometer settings and further functions Moving the cursor up or down Storing of displayed test result Performing Zero Performing Test Displaying date and time user countdown Decimal point ON OFF Mod...

Страница 212: ...hows time and date Press clock key The display shows Either press key to accept the last used user count down or press any number key to start entering a new value The entry comprises two digits each Enter minutes and seconds e g 2 minutes 0 seconds Shift 0 2 0 0 Confirm with key The display shows Start countdown with key After countdown has finished the photometer reverts to the previous display ...

Страница 213: ...ing a method The display shows a selection There are two possibilities to select the required method a enter method number directly e g Shift 8 0 to select Bromine b press arrow key or to select the required method from the displayed list Confirm with key 2 3 2 1 Method Information F1 Use F1 key to switch between the compact and the detailed list for method selection Example Line 1 Method number M...

Страница 214: ...in some methods e g Chlorine The photometer then requires the type of determination Press arrow key or to select the required determi nation Confirm with key 2 3 4 Performing Zero The display shows Prepare a clean vial as described in Method and place the vial in the sample chamber making sure that the marks are aligned Press ZERO key The display shows prepare Zero press ZERO Zero accepted prepare...

Страница 215: ...own with key and proceed as described in the mode description The vial must not be placed in the sample chamber Press TEST key Prepare the water sample as described in the method description and place the vial in the sample chamber The display shows the countdown by pressing the TEST key and the countdown is started automatically After the reaction period is finished the measurement starts automat...

Страница 216: ...played if this method is used the next time If there is the possibility to change the chemical species for a method it is described in the manual The arrows indicate the possible chemical species and are printed below the notes of the method PO4 P P2 O5 2 3 8 Storing results Press STORE key while the test result is displayed The display shows We advise you to enter a numeric code up to 6 places A ...

Страница 217: ... Press F3 key The entire data set is printed with date time Code No method and test result Printing example 100 Chlorine T 0 02 6 mg l Cl2 Profi Mode no 2009 07 01 14 53 09 Test No 1 Code Nr 007 4 80 mg l Cl2 The test No is an internal number that is set automatically if a test result is stored It appears only on the print out 2 3 10 Perform additional measurements To perform additional tests usin...

Страница 218: ...bs 580 nm 940 mAbs 610 nm 950 mAbs 660 nm Select the desired wavelength from the method list or by entering the corresponding method number directly The display shows e g Always carry out zeroing using a filled e g deionised water vial The display shows Carry out measurement of the sample The display shows e g TIP To ensure complete reaction times the user countdown may be helpful chapter 2 2 3 pa...

Страница 219: ...r method list switching off all methods 243 Print 20 Printing all stored results 222 Print code no 22 Print only results of a selected Code No range 224 Print date 21 Print only results of a selected time period 223 Print method 23 Print only results of one selected method 225 Printing parameters 29 Setting of printing options 226 Profi Mode 50 Switching the detailed operator instructions on off 2...

Страница 220: ...he data necessary to run a user concentration method 244 User polynoms 65 Entering the data necessary to run a user polynomial 246 User methods clear 66 Delete all data of a user polynomial or of a concentration method 249 User methods print 67 Print out all data stored with mode 64 concentration or mode 65 polynomial 250 User methods init 69 Initialise the user method system polynomial and concen...

Страница 221: ...tdown even if the key beep is switched off Setting date and time Press MODE Shift 1 2 keys Confirm with key The display shows The entry comprises two digits each Enter year month and day e g 14 May 2009 Shift 0 9 0 5 1 4 Enter hours and minutes e g 3 07 p m Shift 1 5 0 7 Confirm with key Note While confirming date and time with key the seconds are adjusted to zero automatically Key Beep ON 1 OFF 0...

Страница 222: ...ws Press Shift 0 keys to switch the countdown off Press Shift 1 keys to switch the countdown on Confirm with key Notes 1 It is possible to interrupt the working countdown by pressing the key application e g serial analysis The user countdown is also available if the countdown is switched off 2 If the countdown function is switched off the operator is responsible for ensuring the necessary reaction...

Страница 223: ... MODE Shift 1 4 keys Confirm with key The display shows Press Shift 0 keys to switch the signal beep off Press Shift 1 keys to switch the signal beep on Confirm with key Note In the case of methods with reaction periods an acoustic signal still sounds during the last 10 seconds of the countdown even if the key beep is switched off Signal Beep ON 1 OFF 0 0 Mode 4 1 1 ...

Страница 224: ... The display shows Press key for printing out all stored test results The display shows e g After printing the photometer goes back to Mode Menu automatically Note It is possible to cancel the entry by ESC All stored data are printed out See chapter 2 5 1 Data Printing Print Print all Data Start cancel ESC Test No Mode 0 2 ...

Страница 225: ...last day of the required period e g 19 May 2009 Shift 0 9 0 5 1 9 Confirm with key The display shows Press key and all stored results in the selected date range are printed After printing the photometer goes back to mode menu automatically Note It is possible to cancel the entry by ESC If you want to print only results of one day enter the same date twice to determine the period Print sorted date ...

Страница 226: ...ode No e g Shift 1 0 Confirm with key The display shows Press key and all stored results in the selected code number range are printed After printing the photometer goes back to mode menu automatically Note It is possible to cancel the entry by ESC If you want to print only results of one code number enter the same code number twice If you want to print all results without code no code no is 0 ent...

Страница 227: ...irm with key In case of differentiated methods select the required kind of determination and confirm with key The display shows Press key and all stored results of the selected method are printed After printing the photometer goes back to mode menu automatically Note It is possible to cancel the entry by ESC Print 20 Acid demand 30 Alkalinity tot 40 Aluminium T Print method 30 Alkalinity tot Start...

Страница 228: ...ow key or to select the required Protocol Xon Xoff no control Confirm with key Finish with ESC key Flow Control will be set to the selection displayed at is Press Shift 2 keys to select Baud rate The display shows printing parameter 1 Flow control 2 Baud rate cancel ESC Flow Control is Xon Xoff select save cancel ESC Baud rate is 19200 select save cancel ESC 2 Mode 9 2 1 Esc ...

Страница 229: ...ith key The display shows The stored data sets are displayed in chronological order starting with the latest stored test result Press key and all stored results are displayed Press F3 key to print the displayed result Press F2 key to print all results End with ESC Press arrow key to display the following test result Press arrow key to display the previous test result If there are no test results i...

Страница 230: ... g 19 May 2009 Shift 0 9 0 5 1 9 Confirm with key The display shows Press key and all stored results in the selected date range are displayed Press F3 key to print the displayed result Press F2 key to print all selected results End with ESC Note It is possible to cancel the entry by ESC If you want to recall only results of one day enter the same date twice to determine the time period Storage sor...

Страница 231: ...with key The display shows Press key and all stored results in the selected Code No range are displayed Press F3 key to print the displayed result Press F2 key to print all selected results End with ESC Note It is possible to cancel the entry by ESC If you want to recall only results of one code number enter the same code number twice If you want to recall all results without code no code no is 0 ...

Страница 232: ... key In case of differentiated methods select the required kind of determination and confirm with key The display shows Press key and all stored results of the selected method are displayed Press F3 key to print the displayed result Press F2 key to print all selected results End with ESC Storage 20 Acid demand 30 Alkalinity tot 40 Aluminium T Storage method 30 Alkalinity tot Start cancel ESC print...

Страница 233: ...data sets in me mory After pressing keys Shift 1 the following acknow ledgment is displayed Press key to delete ATTENTION All stored test results are deleted or cancel without deleting data by pressing ESC key Note All stored test results are deleted Delete data Delete all data YES 1 NO 0 Delete data Delete data Do not delete ESC 0 Mode 4 3 1 ...

Страница 234: ...amber 5 Pipette 100 ml of water free of calcium to an appropriate beaker note 2 3 6 Add 10 CALCIO H No 1 tablets straight from the foil to the 100 ml of water crush the tablets using a clean stirring rod and dissolve the tablets completely 7 Add 10 CALCIO H No 2 tablets straight from the foil to the same water crush the tablets using a clean stirring rod and dissolve the tablets completely 8 Press...

Страница 235: ...ed or tap water 3 If no water free of Calcium is available these ions can be masked by using EDTA Preparation Add 50 mg a spatula tipful EDTA to 100 ml water and dissolve 4 To achieve the most accurate method blank it is important to adhere exactly to the sample volume of 100 ml Calcium Hardness Method 191 Reset method blank to factory calibration Press MODE Shift 4 0 keys Confirm with key The dis...

Страница 236: ...rom the sample chamber 5 Add exactly 2 ml SPADNS reagent solution to the water sample Caution Vial is filled up to the top 6 Close the vial tightly with the cap and swirl gently several times to mix the contents Calibration M170 Fluoride L Zero deionised water press ZERO Calibration 1 M191 Ca Hardness 2 2 M191 0 Jus Reset 3 M170 Fluoride L Press Shift 0 keys to keep the method blank Press Shift 1 ...

Страница 237: ...ows Confirm with key Back to method selection with ESC key Select Fluoride method with keys Shift 1 7 0 and If an error message appears please repeat adjustment Notes 1 The same batch of SPADNS reagent solution must be used for adjustment and test The adjustment process needs to be performed for each new batch of SPADNS reagent solution see Standard methods 20th 1998 APHA AWWA WEF 4500 F D S 4 82 ...

Страница 238: ...ocedure Return to factory calibration If the user calibration is deleted the factory calibration is automatically activated Remarks The method Fluoride cannot be calibrated with mode 45 since the test requires a calibration related to the batch of the liquid reagent SPADNS mode 40 chapter Fluoride Method 170 Table No Method Recommended range for user calibration 20 Acid demand 1 3 mmol l 35 Alkali...

Страница 239: ...h basic test100 Chlorine free 215 Iodine Calibration with basic test 100 Chlorine free 220 Iron T 0 3 0 7 mg l Fe 222 Iron PP 0 1 2 mg l Fe 223 Iron TPTZ PP 0 3 0 7 mg l Fe 240 Manganese T 1 2 mg l Mn 242 Manganese PP 0 1 0 4 mg l Mn 243 Manganese HR PP 4 6 mg l Mn 250 Molybdate T 5 15 mg l Mo 252 Molybdate HR PP 10 30 mg l Mo 265 Nitrate TT 10 mg l N 270 Nitrite T 0 2 0 3 mg l N 272 Nitrite LR PP...

Страница 240: ... TT 0 3 6 mg l P 326 Phosphate hydr TT 0 3 0 6 mg L P 316 Phosphonate 1 2 mg l PO4 340 Potassium 3 mg l K 350 Silica 0 5 1 5 mg l SiO2 351 Silica LR PP 1 mg l SiO2 352 Silica HR PP 50 mg l SiO2 212 Sodium hypochlorite 8 360 Sulfate PP 50 mg l SO4 355 Sulfate T 50 mg l SO4 365 Sulfide 0 2 0 4 mg l S 370 Sulfite 3 4 mg l SO3 384 Suspended Solids operating range 386 Turbidity operating range 390 Urea...

Страница 241: ...w key once decreases the displayed result Press keys till the displayed result corresponds to the value of the standard Confirm with key to store the new calibration factor Cancel user calibration by pressing ESC key The display shows Now the method name is displayed inverse and the test result is calculated with the new calibration factor 100 Chlorine T 0 02 6 mg l Cl2 0 90 mg l free Cl2 user cal...

Страница 242: ...ument press MODE Shift 4 6 keys and confirm with key The display shows Press Shift 1 keys to delete user calibration Press Shift 0 keys to keep the valid user calibration The instrument goes back to Zero query automatically user calibration 100 Chlorine T 0 02 6 mg l Cl2 clear user calibration YES 1 NO 0 100 Chlorine T 0 02 6 mg l Cl2 prepare ZERO press ZERO Mode 6 4 1 0 ...

Страница 243: ...only a minimum of operator instructions The criteria specified above in d e f are no longer included Press MODE Shift 5 0 keys in succession Confirm with key The display shows Press Shift 0 keys to switch the Profi Mode off Press Shift 1 keys to switch the Profi Mode on The display shows Confirm with key Note Storage of test results is possible When results are stored the display also shows Profi ...

Страница 244: ... is displayed Methods with a point behind the method number will be displayed in the method selection list Methods without a point will not be displayed in the method selection list Press key or to select the required method from the displayed list Switch with F2 key between active and inactive Select next method activate or inactivate it and continue Confirm with key Cancel without storing by pre...

Страница 245: ...ment goes back to mode menu automatically User method list switch all methods off The program structure requires that the method list must have at least one active switched on method For this reason the instrument activates one method automatically Press MODE Shift 6 2 keys Confirm with key The display shows Press Shift 1 keys to display only one method in the method selection list Press Shift 0 k...

Страница 246: ...ge should be within this range to achieve best results 1000 mAbs 1 Abs 1 E displayed Entering a User Concentration Press MODE Shift 6 4 keys Confirm with key Entry Procedure The display shows Enter a method number in the range from 850 to 859 e g Shift 8 5 0 Confirm with key Note if the entered number has already been used to save a concentration the display shows the query Press Shift 0 or ESC ke...

Страница 247: ...e concentration of the first standard e g Shift 0 0 5 One step back with ESC Press F1 key to reset numerical input Confirm with key The display shows Prepare the first standard and press Test key The display shows the input value and the measured absorp tion value Confirm with key Enter the concentration of the second standard e g Shift 0 1 One step back with ESC Press F1 key to reset numerical in...

Страница 248: ... display shows the input value and the measured absorp tion value Confirm with key Note Perform as described above to measure further stan dards The minimum of measured standards is 2 The maximum of measured standards is 14 S1 to S14 If all required standards or the maximum value of 14 stan dards are measured press Store key The display shows The instrument goes back to the mode menu automati call...

Страница 249: ...onent of the coefficient A e g Shift 3 Confirm with key Successively the instrument queries the data for the other coefficients B C D E and F Note If zero 0 is entered for the value of the coefficient the input of the exponent is omitted automatically Confirm every input with key Enter measurement ranges from 2600 to 2600 mAbs Press or key to change between plus and minus sign Enter the values in ...

Страница 250: ...display shows The instrument goes back to the mode menu automati cally Now the polynomial is stored in the instrument and can be recalled by entering its method number or selecting it from the displayed method list TIP Save all your polynomial data in a written form because in case of power outage e g changing the battery all polynomial data will be lost and must be entered again You might want to...

Страница 251: ...Press MODE Shift 6 6 keys Confirm with key The display shows Enter the number of the User Method you want to delete in the range from 800 to 824 or 850 to 859 e g Shift 8 0 0 Confirm with key The query is displayed Press Shift 1 keys to delete the selected User Method Press Shift 0 keys to keep the valid User Method The instrument goes back to mode menu automatically User m clear choose no _______...

Страница 252: ... and concentration methods can be printed out or transferred with HyperTerminal to a PC Press MODE Shift 6 7 keys Confirm with key The display shows Press key to print out the data e g wavelength unit of all stored User Methods The display shows e g After data transfer the photometer goes back to mode menu automatically M800 M803 User m print Start Mode 7 6 ...

Страница 253: ...fined state ATTENTION All stored user methods polynomial concentration are deleted with initialisation Press MODE Shift 6 9 keys Confirm with key The display shows Confirm with key The query is displayed Press Shift 1 keys to start initialisation Press Shift 0 keys to to cancel without initialisation The instrument goes back to mode menu automatically User m init Start Initialising YES 1 NO 0 0 Mo...

Страница 254: ... display shows Enter the temperature value T in the range between 3 and 53 C and confirm with key If F was selected enter the temperature value in the range between 37 and 128 F The display shows Enter the value for Calcium hardness CH in the range between 50 and 1000 mg l CaCO3 and confirm with key The display shows Enter the value for Total Alkalinity TA in the range between 5 and 800 mg l CaCO3...

Страница 255: ...with key and enter a value in the defined range Selection of temperature unit Entering the temperature value is possible in degree Celsius or degree Fahrenheit Therefore the following preselection is once required Press MODE Shift 7 1 keys Confirm with key The display shows Press Shift 1 keys to select degree Celsius Press Shift 2 keys to select degree Fahrenheit The instrument goes back to mode m...

Страница 256: ...display about ten units Confirm with key LCD contrast 2 4 10 Instrument special functions service Photometer Information Press MODE Shift 9 1 keys Confirm with key This method informs you about the current software version about the current detected mains power supply about the number of performed tests and free memory capacity Press arrow key to display the number of performed tests and free memo...

Страница 257: ...the instrument to the serial interface of a computer the optional connection cable with integrated electronic system is required It is possible to update new software applications and additional languages via the internet Please find detailed information at our homepage in the download area as soon as available How to open and close the battery compartment cover see chapter 2 1 3 Please Note To pr...

Страница 258: ...256 MaxiDirect_2c 03 2010 ...

Страница 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...

Страница 260: ...ometer in plastic case 4 batteries Type AA LR 6 1 Instruction manual 1 Guarantee declaration 1 Certificate of compliance Adapter for 16 mm Ø vials Adapter for 13 mm Ø vials Round vials with cap height 48 mm Ø 24 mm Round vials with cap height 90 mm Ø 16 mm Cleaning brush Stirring rod plastic Reagent sets IRIM module and connection cable with integrated electronic system are not part of the standar...

Страница 261: ...nd solvent resistant touch sensitive keyboard with integral beeper as acoustic indicator Power supply 4 batteries Type AA LR 6 lifetime approx 26 hours continuous use or 3500 tests Auto off 20 minutes after last function 30 seconds acoustical signal before switch off Dimensions approx 210 x 95 x 45 mm unit approx 395 x 295 x 106 mm case Weight unit approx 450 g Working condition 5 40 C at max 30 9...

Страница 262: ... KI Potassium iodide KS 4 3 Acid demand to pH 4 3 this method is similar to Total Alkalinity but converted into the unit mmol l as the German DIN 38409 demand TDS Total Dissolved Solids LR Low Range MR Medium Range HR High Range C Reagents from Chemetrics L Liquid reagent P Powder reagent PP Powder Pack T Tablet TT Tube Test DEHA N N Diethylhydroxylamine DPD Diethyl p phenylendiamine DTNB Ellmans ...

Страница 263: ...warning signal every 3 minutes warning signal every 12 seconds warning signal the instrument switches itself off capacity of the battery is too low change the batteries Jus Overrange E4 The user calibration is out of the accepted range Please check the standard reaction time and other possible faults Repeat the user calibration Jus Underrange E4 Overrange E1 The concentration of the standard is to...

Страница 264: ...considering the tolerances of each reading they are the same For this reason the combined Chlorine is most likely zero Example 2 The reading for free Chlorine is under the detection limit The instrument is not able to calculate the combined Chlorine In this case the combined Chlorine is most likely the same as the total Chlorine Example 3 The reading for total Chlorine is exceeding the range The i...

Страница 265: ...is no selection between differentiated free or total Profi Mode is switched on Switch Profi Mode off with Mode 50 The pre programmed countdown is not displayed Countdown is not activated and or the Profi Mode is activated Switch the countdown on with Mode 13 and or switch the Profi Mode off with Mode 50 It seems that a method is not available Method is not activated in the user method list Activat...

Страница 266: ...oduct name MaxiDirect conforms to the following standards EMC Emission EN 61326 1997 A1 1998 A2 2001 A3 2003 Devices for class B Immunity EN 61326 1997 A1 1998 A2 2001 A3 2003 Requirements according to table A1 industrial environment further applicable standards 61000 4 3 1996 A1 2002 The product conforms to the regulations of the directive 2004 108 EG of the European Parliament and the Council of...

Страница 267: ...265 MaxiDirect_2c 03 2010 ...

Страница 268: ...ermany Tintometer AG Hauptstraße 2 5212 Hausen AG Tel 41 0 56 4422829 Fax 41 0 56 4424121 info tintometer ch www tintometer ch Switzerland The Tintometer Limited Lovibond House Solar Way Solstice Park Amesbury SP4 7SZ Tel 44 0 8452 264654 Fax 44 0 1980 625412 sales tintometer com www tintometer com UK Tintometer South East Asia Unit C 11 3A 3 Two Square No 2 Jalan 19 1 Petaling Jaya 46300 Selangor...