1.1 Methods
Oxygen, dissolved
with Vacu-vials
®
K-7553 (see Notes)
10 – 800 μg/l O
2
Insert the adapter for 13 mm Ø round vials.
1. Place the blank in the sample chamber. The blank is part
of the test kit.
2. Press
ZERO
key.
3. Remove the blank from the sample chamber.
4. Water should flow through the special sample container
for several minutes to remove any air bubbles sticking
at the surface.
The water must flow from the bottom to the top.
5. When the sample container is bubble-free press one
Vacu-vial
®
into the lower edge of the sample contai-
ner. The Vacu-vial
®
breaks at the neck and the vial fills
automatically.
A small volume of inert gas remains in the Vacu-vial
®
.
6. Remove the Vacu-vial
®
point downwards from the
sample container immediately.
As the contents of the vial has a higher density than
water, it is important to remove the vial from the sam-
ple container within 5 seconds to prevent any loss of
reagent.
7. The Vacu-vial
®
is closed with one finger (covered with
a glove) to prevent entry of air. Invert the vial several
times. Dry the outside of the vial.
8. Place the Vacu-vial
®
in the sample chamber.
9. Press
TEST
key.
The result is shown in the display in μg/l Oxygen.
9
2
2
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Содержание Maxi Direct
Страница 1: ...Photometer System axi GB Instruction manual ...
Страница 2: ......
Страница 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...
Страница 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 125: ...123 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 Some solids may not dissolve 2 Conversion mg l NO3 mg l N x 4 43 3 N NO3 ...
Страница 143: ...141 MaxiDirect_2c 03 2010 ...
Страница 187: ...185 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 If Sulfate is present a cloudy solution will appear ...
Страница 189: ...187 MaxiDirect_2c 03 2010 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will appear ...
Страница 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 206: ...204 MaxiDirect_2c 03 2010 ...
Страница 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...
Страница 210: ...208 MaxiDirect_2c 03 2010 ...
Страница 258: ...256 MaxiDirect_2c 03 2010 ...
Страница 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...
Страница 267: ...265 MaxiDirect_2c 03 2010 ...