11
MaxiDirect_2c 03/2010
1.1 Methods
The precision of Lovibond
®
Reagent Systems (tablets, powder packs and tube tests) is identical
to the precision specified in standards literature such as American Standards (AWWA), ISO
etc.
Most of the data referred to in these standard methods relates to Standard Solutions.
Therefore they are not readily applicable to drinking-, boiler- or waste-water, since
various interferences can have a major influence on the accuracy of the method.
For this reason we don’t state such potentially misleading data.
Due to the fact that each sample is different, the only way to check the tolerances (‘precision’)
is the Standard Additions Method.
According to this method, first the original sample is tested. Then further samples (2 to 4) are
taken and small amounts of a Standard Solution are added, and further results are obtained.
The amounts added range from approximately half, up to double the amount present in the
sample itself.
These supplementary results make it possible to estimate the actual concentration of the
original sample by comparison.
Literature
The reagent formulations are based on internationally recognised test methods. Some are
described in national and/or international guidelines.
Deutsche Einheitsverfahren zur Wasser-, Abwasser- und Schlammuntersuchung
1.
Standard Methods for the Examination of Water and Wastewater; 18th Edition, 1992
2.
Photometrische Analysenverfahren, Schwedt,
3.
Wissenschaftliche Verlagsgesellschaft mbH, Stuttgart 1989
Photometrische Analyse, Lange / Vejdelek, Verlag Chemie 1980
4.
Colorimetric Chemical Analytical Methods, 9th Edition, London
5.
Notes for searching:
Active Oxygen
->
Oxygen, activ
Alkalinity-m
->
Alkalinity, total
Alkalinity, total
->
Alkalinity, total
Biguanide
->
PHMB
Monochloramine
->
Chloramine, mono
Calcium Hardness
->
Hardness, Calcium
Total Hardness
->
Hardness, total
m-Value
->
Alkalinity, total
p-Value
->
Alkalinity-p
Silicon dioxide
->
Silica
total Alkalinity
->
Alkalinity, total
total Hardness
->
Hardness, total
Langelier Saturation
->
Mode function 70
Index (Water Balance)
Содержание Maxi Direct
Страница 1: ...Photometer System axi GB Instruction manual ...
Страница 2: ......
Страница 9: ...7 MaxiDirect_2c 03 2010 Part 1 Methods ...
Страница 115: ...113 MaxiDirect_2c 03 2010 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Страница 125: ...123 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 Some solids may not dissolve 2 Conversion mg l NO3 mg l N x 4 43 3 N NO3 ...
Страница 143: ...141 MaxiDirect_2c 03 2010 ...
Страница 187: ...185 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 If Sulfate is present a cloudy solution will appear ...
Страница 189: ...187 MaxiDirect_2c 03 2010 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will appear ...
Страница 193: ...191 MaxiDirect_2c 03 2010 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Страница 206: ...204 MaxiDirect_2c 03 2010 ...
Страница 207: ...205 MaxiDirect_2c 03 2010 Part 2 Instrument Manual ...
Страница 210: ...208 MaxiDirect_2c 03 2010 ...
Страница 258: ...256 MaxiDirect_2c 03 2010 ...
Страница 259: ...257 MaxiDirect_2c 03 2010 Part 3 Enclosure ...
Страница 267: ...265 MaxiDirect_2c 03 2010 ...