
Series TID Temperature/Process Indicator
Specifications - Installation and Operating Instructions
Bulletin E-90-TID
The affordable
Series TID
allows user to monitor temperature or a process value.
Temperature ranges are available from -58 to 302°F using one of our PTC or NTC
thermistors. Process values can be displayed from -999 to 999 counts using a 4-20 mA
signal from one of our various transmitters. The process indicator has an adjustable
span and zero on the 4-20 mA models.
Installation
Note:
Unit must be mounted away form vibration, impacts, water, and corrosive gases.
• Cut hole in panel 2.80 x 1.14 in (71 x 29 mm).
• Apply silicone (or rubber gasket) around the perimeter of the hole to prevent leakage.
• Insert unit into hole of panel.
• Slide removable fitting clips onto unit from the back until secure to panel.
• Remove back cover to wire unit.
• Wiring diagram is displayed on the top of the unit.
SPECIFICATIONS
Inputs:
PTC/NTC thermistor or 4-20 mA.
Display:
3-digits; Red, green or blue display.
Range:
-58 to 302°F (thermistor); -999 to 999 counts (4-20 mA).
Power Requirements:
110 VAC, 230 VAC, 12 VAC/DC, 24 VAC/DC.
Accuracy:
1% FS.
Resolution:
1° or 0.01 count.
Front Panel Rating:
IP64.
Agency Approvals:
CE.
INPUT
GND
1
2
3
4
5
7
8
IN
12 VDC
POWER
SUPPLY
CURRENT CONNECTION DIAGRAM
MODEL CODE
TID-
INPUT
1 = THERMISTOR PTC
2 = THERMISTOR NTC
3 = 4-20 mA
1 = 110 VAC
2 = 230 VAC
3 = 12 VAC/DC
4 = 24 VAC/DC
SUPPLY POWER
UNITS
0 = RED
1 = GREEN
2 = BLUE
0 = NONE
1 = °F
2 = °C
DISPLAY
3
[76.2]
1-11/32
[34.13]
2-3/8
[60.38]
1-7/64
[28.18]
THERMISTOR CONNECTION DIAGRAM
PROBE
POWER
1
2
3
4
t°
LOVE CONTROLS
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.
Phone: 219/879-8000
Fax: 219/872-9057
A DIVISION OF
DWYER INSTRUMENTS, INC.
www.lovecontrols.com
e-mail: [email protected]