5.2.2 Changing via LCV-USB-VS
In the advanced mode, the software offers tabs for configuration. The filter frequencies can be adjus-
ted in configuration block STATOR_OPERATION.
5.2.3 Changing via Self-programmed (own) Software
The adjustment for channel A is stored in bytes 13 and 14 of the configuration block “STATOR_OPE-
RATION”. The value for channel B is stored in bytes 15 and 16. Each byte represents a 16-bit value
and corresponds to the cutoff-frequency according following chart:
Please consider that currently adjusted cutoff-frequency will be active after rebooting, only.
For further information regarding programming with own software we refer to document 090110,
Lorenzprotocol “A flexible command set for digital sensors and interfaces”.
6
Pin Assignments
6.1 Pin Assignment for Strain Gauge Sensors with mV Signal
6.2 Pin Assignment for Active Sensors with ±5V or ±10V Signal
Range of values
(decimal)
Value (hexadecimal)
Adjusted cutoff-
frequency
0 ≤ X ≤ 30
0x001E
30 Hz
31 ≤ X ≤ 300
0x012C
300 Hz
301 ≤ X ≤ 1000
0x03E8
1000 Hz
1001 ≤ X ≤ 65536
0x0BB8
3000 Hz
Description
Pin (6 pin socket)
Remark
0V / GND
1
Ground reference for 5V excitation
5V
2
Excitation for 5V strain gauge sensors
Shield
3
Do not connect shield with sensor housing
4
Positive signal input
Signal -
5
Negative signal input
Control signal
6
If available in sensor
Description
Pin (12 pin socket)
Remark
0V / GND
E
Ground reference for 12V supply
12V
F
Supply for 12V active sensors
C
Signal input
Signal -
D
Relating to signal input (internally 0V / GND)
Control signal
K
If available in sensor
Shield
M
Do not connect shield with sensor housing