V-Link
®
-200
User Manual
7.1.1
EXAMPLE: Internal Shunt Calibration
NODE: V-Link-200, 18 bit (262144 A/D values)
CHANNEL TYPE: differential analog input, 0 to 5 V dc input range
SENSOR TYPE: strain gauge, Wheatstone Bridge, full bridge configuration
SENSOR PARAMETERS: application voltage range: +/-2 mV
This is the expected output voltage of the strain gauge based on the range of strain being measured
in the application and the sensitivity of the gauge (volts/strain).
DESIRED OUTPUT: engineering units, microstrain
PROCEDURE:
1. Open SensorConnect and establish communication with the gateway and node. (
).
2. Select Hardware > Input Range for channel 1, /-2 mV from the drop down menu.
3. Under Hardware Offset, select Balance Target for channel 1, select Mid (50%) from the drop
down menu.
4. Select Auto- Balance. When auto- balance is complete, a blue information window will
indicate the balance result.
Figure 41 - Auto-Balance
5. Select Calibration.
6. Select Microstrain from the Unit drop down menu, select the Shunt Cal button enabled on the
right.
57