V-Link
®
-200
User Manual
6.7
Differential Input Channels
The differential measurement channels provide a 5 V dc excitation voltage to the sensor and measure
the resulting sensor signal output. The sensor signal goes through a programmable gain amplifier
(PGA) and is then processed in the node by a18-bit analog-to-digital (A/D) converter over the 5 V dc
range. The resolution of the sensor measurement is dependent on the operating range of the sensor.
If the application is such that only a small portion of the 5 V dc range is being utilized, better resolution
can be achieved by increasing signal amplification and by zeroing the sensor baseline in the
appropriate offset biasing range. Sensor gain and offset values are configurable in SensorConnect.
Figure 34 - Differential Channel Signal Processing
43