![LORAD AFFINITY Series Скачать руководство пользователя страница 204](http://html.mh-extra.com/html/lorad/affinity-series/affinity-series_service-manual_1939281204.webp)
B-2
P/N 9-500-0255
Service Manual
Appendix B: Technical References
Test Points
Table B-2: Power Control Board (PN 1-003-0445) Test Points
Table B-3: C-arm Safety Board (PN 1-003-0440) Test Points
Test Point
Voltage/Signal
Function
TP1
Vref
Active = 10.5 V
TP2
M+
Motor + = +11.7 V compression up; -11.7 V
compression down; Reference to TP3
TP3
M-
Motor - = -11.7 V compression up; +11.7 V
compression down; Reference to TP2
TP5
Vcc
+5 V
TP6
Sense
Active = 200 mV
TP7
MGND
---
TP8
Vbb
+13.6 V
TP9
VM
+20 V
Test Point
Voltage/Signal
Test Point
Voltage/Signal
TP1
Vref
TP11
GND
TP2
Acc0
TP12
-12 V
TP3
Acc1
TP13
Vcc
TP4
Acc2
TP14
+5 V
TP5
Not Used
TP15
-5 V
TP6
ACK OK
TP16
TPY
TP7
DAC 1
TP17
TPX
TP8
DAC 0
TP18
Not Used
TP9
+5 V
TP19
AGND
TP10
+12 V
TP20
DGND
Содержание AFFINITY Series
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ...Service Manual for Affinity Series Mammography Systems Part Number 9 500 0255 Revision 005 ...
Страница 6: ......
Страница 20: ...xiv P N9 500 0255 Service Manual Preface Acronym List ...
Страница 116: ...4 60 P N9 500 0255 Service Manual Chapter 4 Calibrations and Performance Checks Maintenance Procedures ...