Declaration of Conformity

Conditions

The meters are permitted a worst case error of 1% of  A/D  range during electro-magnetic
disturbance, and must recover automatically when disturbance ceases without the need for human
intervention, such as resetting, power-down etc.

Note: RF interference in the range 20MHz to 75MHz, with an amplitude greater than 8V  on the
power cables may cause the display to show “Err” .  This effect will cease when the interference
is removed. This effect may be cured by adding clip-on ferrite cores around the power cable, close
to the display.

The meters covered by this certificate must be installed in adherence to the following conditions :-

Signal cabling shall be routed separately to power carrying cabling (includes relay output wiring)
All signal cabling shall be screened. The screen shall only be terminated to the power earth terminal

This is to confirm that the Product covered by this declaration have been designed and manufactured
to meet the limits of the following  EMC Standard :

EN61326-1:1997

and has been designed to meet the applicable sections of the following safety standards

EN61010-1:2001

Declaration Reference

: EasyReader

Issue Date

: 16 December 2003

Products Covered

: EasyReader series

Title

: DOC-EasyReader

Declared as true and correct, for and on behalf of London Electronics Ltd.

J.R.Lees      Director

17

Содержание EasyReader ER4P

Страница 1: ...ation Rear case screws please note The rear panel is held in place with finger screws which only need to be gently tightened Do not use tools to tighten or loosen the screws as this could cause damage...

Страница 2: ......

Страница 3: ...t position selection 7 8 11 Declaration of Conformity 17 Filter time constant selection 7 8 Front Panel controls 11 General Description 3 Getting Started 4 Introduction 1 Input connection examples 5 L...

Страница 4: ...ent replacement Or we may have no stock at the time you urgently need it You may also need to pay extra carriage charges if you want a fast guaranteed courier service Warranty repairs or replacements...

Страница 5: ...r 110V supply or 1A for DC supplies in the range 11 30VDC Check that the model number and supply voltage suit your application before you install the meter Don t touch any circuitry after you have con...

Страница 6: ...the meter in engineering units to give up to 1999 or 9 999 display range with selectable decimal point position You can either scale the meter theoretically using stored calibration figures or direct...

Страница 7: ...cables through the cable glands Connect the signal and power cables to the appropriate screw terminal connectors See our connection drawing to check that you are using the correct terminals Check bef...

Страница 8: ...arry electrical noise likely to interfere with your measurement Input Connection Examples Loop powered 2 wire 4 20mA transmitter I P 4 20mA from active source Direct 0 10V I P 0 10V from active source...

Страница 9: ...e meter either from a calibrator or a sensor in the system This is known as the DIRECT calibration method Signals must be stable Connect the calibration mode terminal to common to enable this method T...

Страница 10: ...been selected When correct press the OK button 2 How to set ZERO Press the ZERO button for 3 seconds You ll see In LO briefly then a value with one digit brighter than the other This is 4 000 for a 4...

Страница 11: ...ow you to move the decimal point with the UP DOWN buttons 4 When correctly set press OK 2 How to calibrate Zero 1 Apply the Lowest cal signal and ensure it is steady before proceeding 2 Press the ZERO...

Страница 12: ...ESETinput Tare Command Press the front panel tare switch for more than 2 seconds or make a momentary contact closure to the TARE terminal and Common if you want to force the display to 0 Following rea...

Страница 13: ...MODE A contact closure which selects either direct or theoretical calibration method Contact open Theoretical calibration Contact closed Direct calibration TARE Contact closure resets the displayed v...

Страница 14: ...e unlocked ZERO Press for more than 3 seconds Lets you adjust the meter zero calibration SPAN d p Press for more than 3 seconds Lets you adjust the meter span calibration decimal point position filter...

Страница 15: ...eight 102mm A 57mm high model is also available Display Rangability Fully digitally scalable for ZERO and SPAN readings anywhere in the range 1999 to 9999 with selectable decimal point positon Viewing...

Страница 16: ...if the wall material is not suitable to take screws directly Route cables neatly away from the display If the cables may become damaged in the environment protect them with suitable conduit or trunkin...

Страница 17: ...ou may need to use screw plugs or other screw accessories if the material is not suitable to take screws directly Mount the brackets first then fit the display to the brackets Route cables neatly away...

Страница 18: ...Record of Revisions 16 December 2003 Product released 25 March 2004 Added Wiring and Jumper record sheet to page 16 8 November 2005 Added side view for wall mounting case on page 13 15...

Страница 19: ...16 Wiring and jumper record For future reference record all the jumper positions and wiring colours you used in your installation...

Страница 20: ...rs covered by this certificate must be installed in adherence to the following conditions Signal cabling shall be routed separately to power carrying cabling includes relay output wiring All signal ca...

Отзывы: